Head to head comparisons in performance of CD4 point-of-care assays: a Bayesian meta-analysis (2000–2013)
Open Access
- 11 July 2014
- journal article
- Published by ScienceOpen in ScienceOpen Research
Abstract
Timely detection, staging, and treatment initiation are pertinent to controlling HIV infection. CD4+ cell-based point-of-care (POC) devices offer the potential to rapidly stage patients, and decide on initiating treatment, but a comparative evaluation of their performance has not yet been performed. With this in mind, we conducted a systematic review and meta-analyses. For the period January 2000 to April 2014, 19 databases were systematically searched, 6619 citations retrieved, and 25 articles selected. Diagnostic performance was compared across devices (i.e., PIMA, CyFlow, miniPOC, MBioCD4 System) and across specimens (i.e., capillary blood vs. venous blood). A Bayesian approach was used to meta-analyze the data. The primary outcome, the Bland–Altman (BA) mean bias (which represents agreement between cell counts from POC device and flow cytometry), was analyzed with a Bayesian hierarchical normal model. We performed a head-to-head comparison of two POC devices such as PIMA and PointCareNOW CD4. PIMA appears to perform better vs. PointCareNOW with venous samples (BA mean bias: –9.5 cells/μL; 95% CrI: –37.71 to 18.27, vs. 139.3 cells/μL; 95% CrI: –0.85 to 267.4, mean difference = 148.8, 95% CrI: 11.8, 285.8); however, PIMA's best performed when used with capillary samples (BA mean bias: 2.2 cells/μL; 95% CrI: –19.32 to 23.6). Sufficient data were available to allow pooling of sensitivity and specificity data only at the 350 cells/μL cutoff. For PIMA device sensitivity 91.6 (84.7–95.5) and specificity was 94.8 (90.1–97.3), respectively. There were not sufficient data to allow comparisons between any other devices. PIMA device was comparable to flow cytometry. The estimated differences between the CD4+ cell counts of the device and the reference was small and best estimated in capillary blood specimens. As the evidence stands, the PointCareNOW device will need to improve prior to widespread use and more data on MBio and MiniPOC are needed. Findings inform implementation of PIMA and improvements in other CD4 POC device prior to recommending widespread use.Keywords
This publication has 19 references indexed in Scilit:
- Point-of-Care Testing for Infectious Diseases: Diversity, Complexity, and Barriers in Low- And Middle-Income CountriesPLOS Medicine, 2012
- Emerging technologies for point-of-care CD4 T-lymphocyte countingTrends in Biotechnology, 2012
- QUADAS-2: A Revised Tool for the Quality Assessment of Diagnostic Accuracy StudiesAnnals of Internal Medicine, 2011
- Mortality Reduction Associated with HIV/AIDS Care and Antiretroviral Treatment in Rural Malawi: Evidence from Registers, Coffin Sales and FuneralsPLOS ONE, 2010
- Patient retention in antiretroviral therapy programs up to three years on treatment in sub‐Saharan Africa, 2007–2009: systematic reviewTropical Medicine & International Health, 2010
- Loss to Follow-Up of Adults in Public HIV Care Systems in Central Mozambique: Identifying Obstacles to TreatmentJAIDS Journal of Acquired Immune Deficiency Syndromes, 2009
- Clinical applications and availability of CD4+ T cell count testing in sub-Saharan AfricaCytometry Part B: Clinical Cytometry, 2008
- Lymphocyte subsets in healthy children from birth through 18 years of ageJournal of Allergy and Clinical Immunology, 2003
- Extent to which low-level use of antiretroviral treatment could curb the AIDS epidemic in sub-Saharan AfricaThe Lancet, 2000
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986