Load‐related implant reaction of mini‐implants used for orthodontic anchorage
- 19 August 2005
- journal article
- Published by Wiley in Clinical Oral Implants Research
- Vol. 16 (4) , 473-479
Abstract
The purpose of this study was to determine the clinical and biomechanical outcome of two different titanium mini‐implant systems activated with different load regimens. A total of 200 mini‐implants (102 Abso Anchor® and 98 Dual Top®) were placed in the mandible of eight Göttinger minipigs. Two implants each were immediately loaded in opposite direction by various forces (100, 300 or 500 cN) through tension coils. Additionally, three different distances between the neck of the implant and the bone rim (1, 2 and 3 mm) were used. The different load protocols were chosen to evaluate the load‐related implant performance. The load was provided by superelastic tension coils, which are known to develop a virtually constant force. Non‐loaded implants were used as a reference. Following an experimental loading period of 22 and 70 days half of the minipigs were sacrificed, and implant containing bone specimens evaluated for clinical performance and implant stability. Implant loosing was found to be statistically dependent on the tip moment (TM) at the bone rim. Clinical implant loosing were only present when load exceeded 900 cN mm. No movement of implants through the bone was found in the experimental groups, for any applied loads. Over the two experimental periods the non‐loaded implants of one type of implant had a higher stability than those of the loaded implants. Dual Top® implants revealed a slightly higher removal torque compared with Abso Anchor® implants. Based on the results of this study, immediate loading of mini‐implants can be performed without loss of stability when the load‐related biomechanics do not exceed an upper limit of TM at the bone rim.Keywords
This publication has 29 references indexed in Scilit:
- Biological and biomechanical evaluation of bone remodelling and implant stability after using an osteotome techniqueClinical Oral Implants Research, 2004
- The Anchorage Quality of Mini-implants towards Translatory and Extrusive ForcesJournal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie, 2003
- A brief review for orthopedic surgeons: Fatigue damage (microdamage) in bone (its determinants and clinical implications)Journal of Orthopaedic Science, 1998
- Reactions of peri-implant tissues to continuous loading of osseointegrated implantsAmerican Journal of Orthodontics and Dentofacial Orthopedics, 1998
- Changes in Peri‐Implant Tissues Subjected to Orthodontic Forces and Ligature Breakdown in MonkeysJournal of Periodontology, 1998
- Motivation und Erfolgsbeurteilung erwachsener Patienten zur kieferorthopädischen Behandlung-Interpretation einer BefragungJournal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie, 1990
- Studies on the efficacy of implants as orthodontic anchorageAmerican Journal of Orthodontics, 1983
- Three-dimensional Stress Distribution Around a Dental Implant at Different Stages of Interface DevelopmentJournal of Dental Research, 1983
- A 15-year study of osseointegrated implants in the treatment of the edentulous jawInternational Journal of Oral Surgery, 1981
- Osseointegrated Titanium Implants:Requirements for Ensuring a Long-Lasting, Direct Bone-to-Implant Anchorage in ManActa Orthopaedica, 1981