Radiation Dose Reduction for Augmentation Mammography

Authors:

Abstract
OBJECTIVE. Patients who undergo cosmetic augmentation have larger and denser breasts and receive higher radiation doses during mammography than women without implants. In this study we evaluated the dose increase and techniques for dose reduction.SUBJECTS AND METHODS. Mean glandular dose to the breast during screening mammography was measured for 206 women who had undergone breast augmentation. For 13 of these women, mean glandular dose from preoperative mammography also was measured. Effective tube current, peak kilovoltage, and breast thickness were measured, and mean glandular dose was calculated for 1,632 images. Two screen-film combinations and three target-filter combinations were studied.RESULTS. For four-view augmentation mammography with a molybdenum-molybdenum (Mo-Mo) target-filter combination, mean glandular dose was reduced 35%, from 10.7 to 7.0 mGy, by changing the screen-film combination from 100 to 190 speed. For four-view augmentation mammography, mean glandular dose was reduced 24% by cha...