In-situ analysis of Pb isotope ratios using laser ablation MC-ICP-MS: Controls on precision and accuracy and comparison between Faraday cup and ion counting systems
- 26 March 2008
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Journal of Analytical Atomic Spectrometry
- Vol. 23 (7) , 968-975
Abstract
This contribution compares in-situlaser ablation multicollector ICP-MS (LA-MC-ICP-MS) measurements of Pb isotope composition using parallel Faraday cup and ion counter detector array to those made using solely Faraday cups. In the former configuration ion counting is used for the low abundance isotope 204Pb as well as 200Hg and 202Hg isotopes required for correction of isobaric interference of 204Hg. Accurate measurement of the differences in gain between ion counters and Faraday cups, required for determination of 208Pb/204Pb, 207Pb/204Pb and 206Pb/204Pb ratios using ion counting, can be made by analysis of standard glasses to within ±0.3% (2s). For analyses made using Faraday cups or combined Faraday cup and ion counting detector arrays mass bias is effectively controlled by external normalization to 208Pb/206Pb ratios measured in the same standard glasses. Provided that He sweep gas flow rates are kept constant and that adequate time is allowed for stabilization after changes in He flow, mass bias can be controlled to within ±0.05% (2s). The precision of measurements of Pb isotope ratios is strongly dependent on ion beam intensities. Internal precision of ≤ 0.02% (2se) on 208Pb/206Pb and 207Pb/206Pb ratios requires ∑Pb ion beam intensity of > ∼30–50 mV. At beam intensities < 100 mV ∑Pb measurement of 208Pb/204Pb, 207Pb/204Pb and 206Pb/204Pb ratios also benefit significantly from the use of ion counting to measure 204Pb intensity, with an overall improvement of 1–2 orders of magnitude in internal standard error. However, the additional uncertainty associated with calibration of the Faraday cup–ion counter relative gain suggest that use of Faraday cup to measure 204Pb is most appropriate at signal intensities > ∼200 mV, equivalent to ∼100 000 counts per second (cps) 204Pb. Correction for 204Hg interference does not substantially contribute to uncertainty as long as 204Hg comprises less than ∼20% of the total 204 ion beam.Keywords
This publication has 21 references indexed in Scilit:
- Variable initial Pb isotopic compositions of rocks associated with the UG2 chromitite, eastern Bushveld ComplexGeochimica et Cosmochimica Acta, 2007
- LEAD ISOTOPE ANALYSIS OF ARCHAEOLOGICAL METALS BY MULTIPLE‐COLLECTOR INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY*Archaeometry, 2006
- Pb isotopic variability in melt inclusions from the EMI–EMII–HIMU mantle end-members and the role of the oceanic lithosphereEarth and Planetary Science Letters, 2005
- Improved in situ isotope analysis of low-Pb materials using LA-MC-ICP-MS with parallel ion counter and Faraday detectionJournal of Analytical Atomic Spectrometry, 2005
- Lead isotope ratio measurements in geological glasses by laser ablation-sector field-ICP mass spectrometry (LA-SF-ICPMS)International Journal of Mass Spectrometry, 2005
- Precise isotope ratio determination of common Pb using quadrupole LA-ICP-MS with optimized laser sampling conditions and a robust mixed-gas plasmaJournal of Analytical Atomic Spectrometry, 2003
- Lead isotopic disequilibrium between sulfide and plagioclase in the bushveld complex and the chemical evolution of large layered intrusionsGeochimica et Cosmochimica Acta, 2003
- Pb Isotopic Variability in Melt Inclusions from Oceanic Island Basalts, PolynesiaScience, 1998
- Climate and Ocean Dynamics and the Lead Isotopic Records in Pacific Ferromanganese CrustsScience, 1997
- Lead isotope ratio measurement of NIST 610 glass by laser ablation inductively coupled plasma mass spectrometrySpectrochimica Acta Part B: Atomic Spectroscopy, 1993