Effect of 3-tesla magnetic resonance imaging on various pressure programmable shunt valves
Authors:
- Takashi InoueTITakashi Inoue
- Yasutaka KuzuYKYasutaka Kuzu
- Kuniaki OgasawaraKOKuniaki Ogasawara
- Akira OgawaAOAkira Ogawa
- 1 August 2005
- journal article
- Published by Journal of Neurosurgery Publishing Group (JNSPG) in Journal of Neurosurgery
- Vol. 103 (2) , 163-165
Abstract
Object: This study investigated the changes in the valve pressure setting of several magnetic pressure-programmable valves after exposure to a 3-tesla magnetic field. Methods: Five each of four types of pressure-programmable shunt valves were tested: Sophy Polaris, Sophy SM8, Codman-Hakim, and Medtronic Strata. First, the valves were advanced toward the 3-tesla static magnetic field. Second, T 1 -, T 2 -, and diffusion-weighted magnetic resonance (MR) images were generated with a radiofrequency magnetic field. Any changes in the pressure setting were observed by visual inspection with a compass or radiography. The pressure settings were changed after exposure to the static magnetic field in all programmable valves except for the Sophy Polaris. All pressure settings studied were unchanged after exposure to both static and radiofrequency magnetic fields (T 1 -, T 2 -, and diffusion-weighted MR imaging) in the Sophy Polaris. Conclusions: The Sophy Polaris valve allows shunt-dependent patients who need a programmable valve to undergo 3-tesla MR imaging.Keywords
This publication has 8 references indexed in Scilit:
- Pressure changes observed in Codman-Medos programmable valves following magnetic exposure and fillipingChild's Nervous System, 2001
- Transcutaneous Pressure-adjustable Valves and Magnetic Resonance Imaging: An Ex Vivo Examination of the Codman-Medos Programmable Valve and the Sophy Adjustable Pressure ValveNeurosurgery, 1997
- Clinical experience with a new pressure-adjustable shunt valveActa Neurochirurgica, 1995
- Transcutaneous pressure adjustable valve for the treatment of hydrocephalus and arachnoid cysts in adultsActa Neurochirurgica, 1993
- Safety of the adjustable pressure ventricular valve in magnetic resonance imaging: problems and solutionsNeuroradiology, 1992
- Do We Need Variable Pressure Shunts?British Journal Of Neurosurgery, 1992
- Clinical experience with a pressure-adjustable valve SOPHY in the management of hydrocephalusChild's Nervous System, 1990
- Increased Intraventricular Pressure without Ventriculomegaly in Children with ShuntsNeurosurgery, 1979