Cone Beam X-ray Luminescence Computed Tomography Based on Bayesian Method
- 26 August 2016
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Medical Imaging
- Vol. 36 (1) , 225-235
Abstract
X-ray luminescence computed tomography (XLCT), which aims to achieve molecular and functional imaging by X-rays, has recently been proposed as a new imaging modality. Combining the principles of X-ray excitation of luminescence-based probes and optical signal detection, XLCT naturally fuses functional and anatomical images and provides complementary information for a wide range of applications in biomedical research. In order to improve the data acquisition efficiency of previously developed narrow-beam XLCT, a cone beam XLCT (CB-XLCT) mode is adopted here to take advantage of the useful geometric features of cone beam excitation. Practically, a major hurdle in using cone beam X-ray for XLCT is that the inverse problem here is seriously ill-conditioned, hindering us to achieve good image quality. In this paper, we propose a novel Bayesian method to tackle the bottleneck in CB-XLCT reconstruction. The method utilizes a local regularization strategy based on Gaussian Markov random field to mitigate the ill-conditioness of CB-XLCT. An alternating optimization scheme is then used to automatically calculate all the unknown hyperparameters while an iterative coordinate descent algorithm is adopted to reconstruct the image with a voxel-based closed-form solution. Results of numerical simulations and mouse experiments show that the self-adaptive Bayesian method significantly improves the CB-XLCT image quality as compared with conventional methods.Keywords
Funding Information
- National Natural Science Foundation of China (No 61601019, 81571836, 61322101)
- Beijing Natural Science Foundation (No. 7164270)
- Fundamental Research Funds for the Central Universities (No. 2015JBM026)
- China Postdoctoral Science Foundation (No. 2015M570032)
- China Scholarship Council (No. 201507095005)
- Guangdong Innovation Team of Image-Guided Therapy (No. 2011S013)
- National Institutes of Health (No. 1R01 EB016777)
This publication has 43 references indexed in Scilit:
- JointL1and total variation regularization for fluorescence molecular tomographyPhysics in Medicine & Biology, 2012
- MAP estimation with structural priors for fluorescence molecular tomographyPhysics in Medicine & Biology, 2012
- Limited-angle x-ray luminescence tomography: methodology and feasibility studyPhysics in Medicine & Biology, 2011
- Spectrally resolving and scattering-compensated x-ray luminescence∕fluorescence computed tomographyJournal of Biomedical Optics, 2011
- Facile Synthesis of Amine-Functionalized Eu3+-Doped La(OH)3 Nanophosphors for BioimagingDiscover Nano, 2010
- Hybrid x‐ray/optical luminescence imaging: Characterization of experimental conditionsMedical Physics, 2010
- Cone Beam Micro‐CT System for Small Animal Imaging and Performance EvaluationInternational Journal of Biomedical Imaging, 2009
- A generalized Gaussian image model for edge-preserving MAP estimationIEEE Transactions on Image Processing, 1993
- Statistic-based MAP image-reconstruction from Poisson data using Gibbs priorsIEEE Transactions on Signal Processing, 1992
- Practical cone-beam algorithmJournal of the Optical Society of America A, 1984