CO2 Laser Welding of Corneal Cuts with Albumin Solder Using Radiometric Temperature Control
Authors:
- Eitan LivnyELEitan Livny
- Nino LoyaNLNino Loya
- Noam KarivNKNoam Kariv
- Avi RavidARAvi Ravid
- Abraham KatzirAKAbraham Katzir
- Dan D. GatonDGDan D. Gaton
- 1 September 2013
- journal article
- Published by S. Karger AG in Ophthalmic Research
- Vol. 50 (3) , 174-179
Abstract
Purpose: To examine the efficacy and reproducibility of CO2 laser soldering of corneal cuts using real-time infrared fiber-optic radiometric control of tissue temperature in bovine eyes (in vitro) and to evaluate the duration of this procedure in rabbit eyes (in vivo). Methods: In vitro experiment: a 6-mm central perforating cut was induced in 40 fresh bovine eyes and sealed with a CO2 laser, with or without albumin soldering, following placement of a single approximating nylon suture. A fiber-optic radiometric temperature control system for the CO2 laser was used. Leaking pressure and histological findings were analyzed and compared between groups. In vivo experiment: following creation of a central perforation, 6 rabbit eyes were treated with a CO2 laser with albumin solder and 6 rabbit eyes were treated with 10-0 nylon sutures. The amount of time needed for completion of the procedures was compared. Results: In vitro experiment: effective sealing was achieved by CO2 laser soldering. Mean (± SD) leaking pressure was 109 ± 30 mm Hg in the bovine corneas treated by the laser with albumin solder compared to 51 ± 7 mm Hg in the sutured control eyes (n = 10 each; p < 0.001). Mean leaking pressures were much lower in the corneal cuts sealed only with the laser without albumin solder (48 ± 12 mm Hg) and in the cuts sealed only with albumin without laser welding (6.3 ± 4 mm Hg) than in the cuts treated with laser welding and albumin solder. In vivo experiment: mean surgical time was 140 ± 17 s in the laser-treated rabbits compared to 330 ± 30 s in the sutured controls (n = 6; p < 0.001). A histopathological study of the rabbit corneas 1 day after laser soldering revealed sealed corneal edges with a small gap bridged by coagulated albumin. The inflammatory reaction was minimal in contrast to the sutured controls. No thermal damage was detected at the wound edges. Conclusions: CO2 laser soldering combined with the fiber-optic radiometer is an effective, reliable, and rapid tool for the closure of corneal wounds, and holds advantages over conventional suturing in terms of leaking pressure and surgical time.Keywords
This publication has 25 references indexed in Scilit:
- Robotic laser tissue welding of sclera using chitosan filmsLasers in Surgery and Medicine, 2009
- Microscopic characterization of collagen modifications induced by low‐temperature diode‐laser welding of corneal tissueLasers in Surgery and Medicine, 2007
- Carbon dioxide laser‐assisted nerve repair: Effect of solder and suture material on nerve regeneration in rat sciatic nerveMicrosurgery, 2003
- Thermal feedback control techniques for transistor–transistor logic triggered CO_2 laser used for irradiation of biological tissue utilizing infrared fiber-optic radiometryApplied Optics, 1994
- Changes in type I collagen following laser weldingLasers in Surgery and Medicine, 1992
- Sutureless End‐to‐End bowel anastomosis using Nd:YAG and water‐soluble intralurninal stentLasers in Surgery and Medicine, 1990
- Techniques for better suturingBritish Journal of Surgery, 1988
- CO2 and argon laser vascular welding: Acute histologic and thermodynamic comparisonLasers in Surgery and Medicine, 1988
- Mechanism of tissue fusion in argon laser‐welded vein‐artery anastomosesLasers in Surgery and Medicine, 1988
- SUTURELESS EXTRA-INTRACRANIAL ANASTOMOSIS BY LASERThe Lancet, 1984