An Inexpensive Flow‐Through Cell and Measurement System for Monitoring Selected Chemical Parameters in Ground Water
Authors:
- Edward E. GarskeEGEdward E. Garske
- Michael R. SchockMSMichael R. Schock
- 1 September 1986
- journal article
- Published by Wiley in Groundwater Monitoring & Remediation
- Vol. 6 (3) , 79-84
Abstract
An inexpensive flow‐through cell was designed and constructed to enable the accurate measurement of redox potential (Eh), conductivity and pH from shallow monitoring wells. The cell accepts sample water from inert gas‐operated bladder pumps. The electrodes are monitored by meters mounted in a portable rugged waterproof carrying case. The cell is useful for monitoring the purging of stagnant water from the well prior to sample collection for chemical analysis, as well as to provide pH, Eh and conductivity data under conditions as close to in situ as presently practical. In conjunction with the development of the flow‐through cell and measuring system, a portable field kit was constructed and equipped to determine the alkalinity of the water via modified Gran plot titration, immediately upon sample collection. In addition to the measurement cell system, an in‐line membrane filtration system was developed, that included valves to sequentially switch to different filter units, in case of membrane plugging.Keywords
This publication has 10 references indexed in Scilit:
- Sampling Tubing Effects on Groundwater SamplesAnalytical Chemistry, 1985
- A Laboratory Evaluation of Ground Water Sampling MechanismsGroundwater Monitoring & Remediation, 1984
- Plumbosolvency reduction by high pH and low carbonate—solubility relationshipsJournal AWWA, 1983
- Guidelines and techniques for obtaining water samples that accurately represent the water chemistry of an aquiferOpen-File Report, 1982
- Adsorption of trace metals during filtration of potable water samples with particular reference to the determination of filtrable lead concentrationThe Analyst, 1981
- Filtration des solutions aqueuses en vue du dosage des cations metalliquesEnvironmental Technology Letters, 1980
- Trace metal ion filtration at pH 5 and 7Analytical Chemistry, 1979
- Thermochemical redox equilibria of ZoBell's solutionGeochimica et Cosmochimica Acta, 1977
- Guidelines for collection and field analysis of ground-water samples for selected unstable constituents1976
- TEMPERATURE CORRECTION IN CONDUCTIVITY MEASUREMENTS1Limnology and Oceanography, 1962