Fast imaging using subencoding data sets from multiple detectors
Authors:
- 1 July 1993
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 30 (1) , 142-145
Abstract
A new fast imaging method using a subencoding data acquisition scheme and a multiple coil receiver system is proposed and demonstrated. In this method, a set of aliased images are produced from receiver coils by using the subencoded data without sacrificing the desired resolution, and resolved to an aliasing‐free image by using the distance‐dependent sensitivity information of each coil. The reduction rate of data acquisition time is proportional to the number of receiver coils. This method can be easily adapted to conventional imaging methods including fast imaging to further reduce the scan time.Keywords
This publication has 10 references indexed in Scilit:
- A decoupled coil detector array for fast image acquisition in magnetic resonance imagingMedical Physics, 1991
- 0.3‐second flash mri of the human heartMagnetic Resonance in Medicine, 1990
- An analysis of fast imaging sequences with steady‐state transverse magnetization refocusingMagnetic Resonance in Medicine, 1988
- Fast MRI data acquisition using multiple detectorsMagnetic Resonance in Medicine, 1988
- An algorithm for NMR imaging reconstruction based on multiple RF receiver coilsJournal of Magnetic Resonance (1969), 1987
- Steady-State Magnetizations in Rapid NMR Imaging Using Small Flip Angles and Short Repetition IntervalsIEEE Transactions on Medical Imaging, 1987
- Rapid Fourier imaging using steady‐state free precessionMagnetic Resonance in Medicine, 1987
- FLASH imaging. Rapid NMR imaging using low flip-angle pulsesJournal of Magnetic Resonance (1969), 1986
- High-Speed Spiral-Scan Echo Planar NMR Imaging-IIEEE Transactions on Medical Imaging, 1986
- Multi-planar image formation using NMR spin echoesJournal of Physics C: Solid State Physics, 1977