A comparison of patient position displacements from body surface laser scanning and cone beam CT bone registrations for radiotherapy of pelvic targets
- 20 June 2013
- journal article
- research article
- Published by Medical Journals Sweden AB in Acta Oncologica
- Vol. 53 (2) , 268-277
Abstract
Background and purpose. Optical surface detection has attractive features as a mean in radiotherapy for patient positioning tasks such as set-up, monitoring and gating. To aid in hitting radiotherapy targets the correlation between detected surface displacements and internal structure displacements is crucial. In this study, we compare set-up displacements derived from a body surface laser scanning (BSLS) system to displacements derived from bone registrations with a cone beam computed tomography (CBCT) system in order to quantify the accuracy and applicability of BSLS for fractionated treatments in the pelvic region. Material and methods. Displacements from concurrent BSLS and CBCT registrations were compared for 40 patients treated in the pelvic region for a total of 170 set-ups. Surface data captured by BSLS at the first treatment fraction (BSLSref) was used as main reference for the BSLS system, while bony structures from the planning CT were used as a reference for the CBCT method. As comparison, the patient outline extracted from the planning CT was used as BSLS reference (CTref). The displacements detected by the CBCT system (skin-marks-only) was also used for comparison. Results. The mean differences (± 1 SD) between the BSLS and CBCT displacements were −0.01 (± 0.17) cm, 0.00 (± 0.21) cm and 0.01 (± 0.17) cm in the lateral, longitudinal and vertical directions, respectively. The median length of the difference was 0.26 cm (0.24–0.29 cm, 95% CI). The median of the difference between CBCT and BSLS displacements based on CTref was 0.37 cm (0.30–0.39 cm) and the median for skin-marks-only was 0.38 cm (0.34–0.42 cm). Conclusions. The BSLS system is a good supplement to the CBCT system for accurate set-up for fractions when no CBCT is deemed necessary for pelvic targets. Inter-fractional skin movement in relation to bone was estimated to be 0.2 cm in the lateral (X), longitudinal (Y) and vertical direction (Z), respectively.Keywords
This publication has 26 references indexed in Scilit:
- A phantom evaluation of Sentinel™, a commercial laser/camera surface imaging system for patient setup verification in radiotherapyMedical Physics, 2012
- Technical performance of a commercial laser surface scanning system for patient setup correction in radiotherapyPhysica Medica, 2011
- Frame-less and mask-less cranial stereotactic radiosurgery: a feasibility studyPhysics in Medicine & Biology, 2010
- Reproducibility of patient setup by surface image registration system in conformal radiotherapy of prostate cancerRadiation Oncology, 2009
- Accuracy of a commercial optical 3D surface imaging system for realignment of patients for radiotherapy of the thoraxPhysics in Medicine & Biology, 2007
- Positioning accuracy of cone-beam computed tomography in combination with a HexaPOD robot treatment tableInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2007
- Clinical experience with a 3D surface patient setup system for alignment of partial-breast irradiation patientsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2006
- Review of 20 years of range sensor developmentJournal of Electronic Imaging, 2004
- A method for registration of 3-D shapesIEEE Transactions on Pattern Analysis and Machine Intelligence, 1992
- Hierarchical chamfer matching: a parametric edge matching algorithmIEEE Transactions on Pattern Analysis and Machine Intelligence, 1988