A randomized controlled comparison of the Bonfils fiberscope and the GlideScope Cobalt AVL video laryngoscope for visualization of the larynx and intubation of the trachea in infants and small children with normal airways
Authors:
- 18 March 2013
- journal article
- research article
- Published by Wiley in Pediatric Anesthesia
- Vol. 23 (10) , 913-919
Abstract
Summary: Background: The Bonfils fiberscope (BF) used without the assistance of a laryngoscope failed to improve the view of direct laryngoscopy in children with normal airways. We hypothesized that if BF is supported by a laryngoscope—as recommended by its inventor—it can provide comparably good visualization of the glottis as the GlideScope® Cobalt AVL video laryngoscope (GS).Methods: We included 100 children with normal airways in a randomized controlled trial. The study consisted of assessing the airway by direct laryngoscopy (DL), followed by intubation using either the BF or the GlideScope. Main outcome measures were the quality of visualization of the larynx by the percentage of glottis opening seen (POGO) and the time needed for intubation of the trachea.Results: Visualization of the larynx (POGO) using the BF was significantly better than with DL (P = 0.016) or with GS (P = 0.001). The DL provided an allover better visualization than GS, although this difference was not significant and solely attributable to children weighing P < 0.001).Conclusion: The Bonfils fiberscope significantly improved the view on the larynx compared with direct laryngoscopy and the GlideScope and enables shorter intubation time than with the GlideScope.Keywords
This publication has 18 references indexed in Scilit:
- Bonfils intubating fibrescope in normal paediatric airwaysBritish Journal of Anaesthesia, 2010
- A comparison of bonfils fiberscope-assisted laryngoscopy and standard direct laryngoscopy in simulated difficult pediatric intubation: a manikin studyPediatric Anesthesia, 2010
- A comparison of the STORZ video laryngoscope and standard direct laryngoscopy for intubation in the Pediatric airway – a randomized clinical trialPediatric Anesthesia, 2009
- Evaluation of the pediatric bonfils fiberscope for elective endotracheal intubationPediatric Anesthesia, 2008
- GlideScope® video laryngoscope: a randomized clinical trial in 203 paediatric patientsBritish Journal of Anaesthesia, 2008
- Early clinical experience with a new videolaryngoscope (GlideScope®) in 728 patientsCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 2005
- A comparison of the intubating laryngeal mask airway and the Bonfils intubation fibrescope in patients with predicted difficult airways*Anaesthesia, 2004
- Assessment of laryngeal view: Percentage of glottic opening scorevs Cormack and Lehane gradingCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1999
- Test statistics and sample size formulae for comparative binomial trials with null hypothesis of non‐zero risk difference or non‐unity relative riskStatistics in Medicine, 1990
- DIRECT LARYNGOSCOPY AND TRACHEAL INTUBATIONThe Lancet, 1944