A Fast, Validated HPLC Method Applied to the Dissolution Test of Gastro‐Resistant Capsules of Pantoprazole Pellets

Authors:

Abstract
A fast and simple liquid chromatographic method has been developed for determination of pantoprazole in delayed release gastro‐resistant capsules employing a reverse‐phase column, mobile phase consisting of acetonitrile‐water (75:25), flow rate 1.0 mL min−1 and 290 nm detection. Selectivity, accuracy, linearity, intra‐ and inter‐day precision were evaluated. The method is not affected by the pellets excipients or induced acid degradation products. Linearity ranged from 2.0–18.0 µg mL−1. The relative standard deviation (RSD) for intra‐day and inter‐day precision varied from 0.92–2.00%. Capsules analysis resulted in 93.48–105.15% of the labeled pantoprazole amount. Recoveries after spiking pellets with pantoprazole ranged from 96.27–102.87%.