Titanium Aneurysm Clips: Part I-Mechanical, Radiological, and Biocompatibility Testing
Authors:
- 1 June 1996
- journal article
- research article
- Published by Wolters Kluwer Health in Neurosurgery
- Vol. 38 (6) , 1158-1164
Abstract
Most aneurysm clips are made of cobalt-based alloys. Although these clips are nonferromagnetic, they still produce artifact that degrades the quality of magnetic resonance (MR) images. A new aneurysm clip of pure titanium was developed to minimize artifact on postoperative MR images. We evaluated these clips in a series of mechanical tests in vitro, biocompatibility tests in rabbits, and radiological tests in greyhound dogs. The clip sizes and shapes matched those of conventional aneurysm clips. The average closing forces ranged between 151.6 and 181.8 g and were not diminished by repeated sterilization or stress. After > 20 million cycles of high-pressure and high-frequency pulsations, the clips did not open and the closing forces were not reduced. Titanium aneurysm clips implanted in the subarachnoid space of 12 rabbits for 1 or 6 months produced mild gliosis identical to that produced by implantation of cobalt alloy clips in 12 control rabbits. Based on pre- and postoperative weights and electron microscopic scans, the titanium implants did not corrode. The artifact on computed tomographic and MR imaging produced by a titanium clip placed on the internal carotid artery of a greyhound was less than that produced by an identical cobalt-chrome alloy clip by a factor of two to three. This study demonstrated that titanium aneurysm clips are mechanically equivalent to conventional clips, biocompatible, and corrosion resistant. Furthermore, titanium clips have superior imaging characteristics, creating less computed tomographic and MR imaging artifact and permitting better resolution of anatomic structures than cobalt alloy clips.Keywords
This publication has 21 references indexed in Scilit:
- MR procedures and biomedical implants, materials, and devices: 1993 update.Radiology, 1993
- Placement of a ferromagnetic intracerebral aneurysm clip in a magnetic field with a fatal outcome.Radiology, 1993
- High-field-strength MR imaging and metallic biomedical implants: an ex vivo evaluation of deflection forcesAmerican Journal of Roentgenology, 1988
- Aneurysm Clip Motion during Magnetic Resonance Imaging: In Vivo Experimental Study with Metallurgical Factor AnalysisNeurosurgery, 1985
- Transmigration of titanium dioxide (TiO2) particles in rats after inhalation exposureExperimental and Molecular Pathology, 1985
- Pulmonary response of rats exposed to titanium dioxide (TiO2) by inhalation for two yearsToxicology and Applied Pharmacology, 1985
- The effects of magnetic resonance imaging on different types of microsurgical clips.Journal of Neurology, Neurosurgery & Psychiatry, 1985
- Tissue compatibility of different intracranial implant materials: In-vivo and in-vitro studiesActa Neurochirurgica, 1984
- Potential hazards and artifacts of ferromagnetic and nonferromagnetic surgical and dental materials and devices in nuclear magnetic resonance imaging.Radiology, 1983
- Transorbital Approach to the Middle Cerebral Artery of the Squirrel Monkey: A Technique for Experimental Cerebral Infarction Applicable to Ultrastructural StudiesStroke, 1970