Quantitative Dimension Measurement of ICF Components Using Xradia MicroXCT Microscope

Authors:

Abstract
National Ignition Facility (NIF) specifications have stringent dimensional accuracy requirements on target components. For example, the laser-hole diameter on an ablator capsule must be characterized to ±0.5 μm to ensure proper fill tube insertion and to minimize the glue joint mass to <2.5 ng. A charge-coupled-device-based X-ray radiography and tomography instrument (commercially obtained from Xradia, Inc.) is used in target metrology where sample opacity precludes the use of optical techniques; however, the built-in caliper for dimensional measurement cannot provide the required accuracy. The instrument has three main error sources: (a) point projection magnification, (b) imaging lens distortion, and (c) phase contrast shift. The sample feature size dictates the calibration strategy. For large features such as the shell diameter, (a) and (b) dominate the error budget. The built-in caliper is accurate to ~2 to 3%, corresponding to a ±50-μm error for a 2000-μm NIF capsule. In this work, we develop...

This publication has 11 references indexed in Scilit: