A Patient-Specific Machine Learning based EEG Processor for Accurate Estimation of Depth of Anesthesia
Authors:
- 1 October 2018
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 1-4
Abstract
An electroencephalograph (EEG) based classification processor for the depth of Anesthesia (DoA) during the intraoperative procedure is presented. To enable a DoA to monitor the correct estimation across a range of patients, a novel feature extraction along with machine learning processor is utilized. The decisions are solely based on seven features extracted from EEG along with the EMG signal for motion artifacts rejection. To extract the features efficiently on hardware, a 128-point FFT is proposed that achieves an area reduction and energy/FFT-operation by 39% and 58%, respectively, compared to the conventional. A simple decision tree is used to perform a multiclass DoA classification. The system is synthesized using a 65nm process and experimental verification is done using FPGA based on the subset of patients from the University of Queensland Vital Signs. The proposed patient-specific DoA classification processor achieves a classification accuracy of 79%.Keywords
This publication has 11 references indexed in Scilit:
- An EEG-NIRS Multimodal SoC for Accurate Anesthesia Depth MonitoringIEEE Journal of Solid-State Circuits, 2018
- A 0.21 μJ patient-specific REM/Non-REM sleep classifier for Alzheimer patients2017
- A Pseudo OFDM With Miniaturized FSK Demodulation Body-Coupled Communication Transceiver for Binaural Hearing Aids in 65 nm CMOSIEEE Journal of Solid-State Circuits, 2017
- A high accuracy and low latency patient-specific wearable fall detection system2017
- Clinical Electroencephalography for AnesthesiologistsAnesthesiology, 2015
- Age-dependent electroencephalogram (EEG) patterns during sevoflurane general anesthesia in infantseLife, 2015
- University of Queensland Vital Signs DatasetAnesthesia & Analgesia, 2012
- General Anesthesia, Sleep, and ComaNew England Journal of Medicine, 2010
- Monitoring the Depth of AnaesthesiaSensors, 2010
- The Relationship Between Bispectral Index and Electroencephalographic Parameters During Isoflurane AnesthesiaAnesthesia & Analgesia, 2004