Influence of cochlear implantation on peripheral vestibular receptor function
- 1 June 2010
- journal article
- Published by Wiley in Otolaryngology–Head and Neck Surgery
- Vol. 142 (6) , 809-813
Abstract
Objective: The objectives of this study were 1) to assess the influence of a cochlear implantation on peripheral vestibular receptor function in the inner ear in the implant and in the nonimplant side, and 2) to analyze a possible correlation with resulting vertigo symptoms.Study Design: Prospective clinical study.Setting: Cochlear implant center at tertiary referral hospital.Subjects and Methods: A total of 32 patients, aged 15 to 83 years, undergoing cochlear implantation were assessed pre‐ and postoperatively for caloric horizontal semicircular canal response and vestibular‐evoked myogenic potentials of the sacculus, and postoperatively for subjective vertigo symptoms. Patients with vertigo were compared with patients without symptoms with regard to the findings of the vestibular function tests.Results: Cochlear implantation represents a significant risk factor for horizontal semicircular canal impairment (P < 0.001) and sacculus impairment (P = 0.047) in the implanted ear. In eight of 16 patients with preoperatively preserved caloric response, this response was decreased postoperatively. Before surgery, 14 of 30 patients had regular vestibular‐evoked myogenic responses. Two months after implantation, six patients had a new loss and another six showed depression of sacculus function on vestibular‐evoked myogenic potentials testing. The impaired vestibular function did not correlate with vertigo symptoms. Function on the contralateral side remains unaffected (P > 0.05).Conclusion: Cochlear implantation is a relevant risk factor for damage of peripheral vestibular receptor function. Therefore, preservation not only of residual hearing function but also of vestibular function should be aimed for, by using minimally invasive surgical techniques.Keywords
This publication has 18 references indexed in Scilit:
- Prevalence and Characteristics of Preoperative Balance Disorders in Cochlear Implant CandidatesAnnals of Otology, Rhinology & Laryngology, 2008
- Incidence and quality of vertigo symptoms after cochlear implantationThe Journal of Laryngology & Otology, 2008
- Does the surgical approach in cochlear implantation influence the occurrence of postoperative vertigo?Otolaryngology–Head and Neck Surgery, 2008
- Cochlear Implant and Vestibular FunctionOtology & Neurotology, 2006
- Histopathology of the Peripheral Vestibular System after Cochlear Implantation in the HumanOtology & Neurotology, 2006
- Vestibular-evoked myogenic potentials in cochlear implant childrenActa Oto-Laryngologica, 2006
- Normative data for P1/N1-latencies of vestibular evoked myogenic potentials induced by air- or bone-conducted tone burstsClinical Neurophysiology, 2005
- Vestibular Effects of Cochlear ImplantationThe Laryngoscope, 2004
- Vestibular Dysfunction after Cochlear ImplantationOtology & Neurotology, 2003
- Histopathologic Changes in the Vestibule after Cochlear ImplantationOtolaryngology–Head and Neck Surgery, 2002