Vecuronium- and Esmolol-Induced Pseudohypernatremia Due to Drug Interference With Ion-Selective Electrodes
Open Access
- 1 January 2020
- journal article
- research article
- Published by Wolters Kluwer Health in Critical Care Explorations
- Vol. 2 (1) , e0073
Abstract
Objectives: We observed that patients treated with continuous vecuronium or esmolol infusions showed elevated plasma sodium measurements when measured by the routine chemistry analyzer as part of the basic metabolic panel (Vitros 5600; Ortho Clinical Diagnostics, Raritan, NJ), but not by blood gas analyzers (RAPIDLab 1265; Siemens, Tarrytown, NY). Both instruments use direct ion-selective electrode technology, albeit with different sodium ionophores (basic metabolic panel: methyl monensin, blood gas: glass). We questioned if the basic metabolic panel hypernatremia represents artefactual pseudohypernatremia. Design: We added vecuronium bromide or esmolol hydrochloric acid to pooled plasma samples and compared sodium values measured by both methodologies. We queried sodium results from the electronic medical records of patients admitted at Children’s Hospital of Philadelphia from 2016 to 2018 and received vecuronium and/or esmolol infusion treatment during their admissions. Setting: PICU of a quaternary, free-standing children’s hospital. Patients: Children admitted to the hospital who received vecuronium and/or esmolol infusion. Measurements and Main Results: Sodium was measured in pooled plasma samples by basic metabolic panel and blood gas methodologies after adding vecuronium bromide or esmolol hydrochloric acid, leading to a dose-response increase in basic metabolic panel sodium measurements. A repeated measures regression analysis of our electronic medical records showed that the vecuronium dose predicted the Δ sodium (basic metabolic panel–blood gas) sodium within 12 hours of the vecuronium administration (p < 0.0018). Esmolol showed a similar trend (p = 0.13). This occurred primarily in central line samples with continuous vecuronium or esmolol infusions. Conclusions: Vecuronium and esmolol can falsely elevate direct ion-selective electrode sodium measurements on Vitros chemistry analyzers. Unexpectedly high sodium measurements in patients receiving vecuronium and/or esmolol infusions should be further investigated with an alternate sample type (i.e., peripheral blood) or measurement methodology (i.e., blood gas) to guide treatment decisions.Keywords
This publication has 18 references indexed in Scilit:
- Prevention of Iatrogenic Anemia in Critical and Neonatal CareAdvances in Clinical and Experimental Medicine, 2015
- Disorders of Plasma Sodium — Causes, Consequences, and CorrectionNew England Journal of Medicine, 2015
- Clinical review: Practical approach to hyponatraemia and hypernatraemia in critically ill patientsCritical Care, 2012
- Analytical evaluation of the VITROS® 5600 Integrated System in a pediatric setting and determination of pediatric reference intervalsClinical Biochemistry, 2010
- Ion Selective Electrodes (ISEs) and interferences—A reviewClinica Chimica Acta, 2010
- Reduction in Red Blood Cell Transfusions Among Preterm Infants: Results of a Randomized Trial With an In-Line Blood Gas and Chemistry MonitorPediatrics, 2005
- A Comparison of the Measurement of Sodium and Potassium by Flame Photometry and Ion-Selective ElectrodeAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 1985
- Oversmoothed Nonparametric Density EstimatesJournal of the American Statistical Association, 1985
- Flame Photometry A Rapid Analytical ProcedureIndustrial & Engineering Chemistry Analytical Edition, 1945
- A SPECIFIC REAGENT FOR THE RAPID GRAVIMETRIC DETERMINATION OF SODIUMJournal of the American Chemical Society, 1928