Intraocular pressure measurement by three different tonometers in primary congenital glaucoma
Open Access
- 1 January 2015
- journal article
- Published by Knowledge E DMCC in Journal of Ophthalmic and Vision Research
- Vol. 10 (1) , 43-8
Abstract
To determine the agreement between intraocular pressure (IOP) measurements using an automated non-contact tonometer (NCT), Goldmann applanation tonometer (GAT), and the ocular response analyzer (ORA) in subjects with primary congenital glaucoma (PCG). Twenty-nine eyes of 17 PCG patients underwent IOP measurements using NCT, GAT and ORA. Variables obtained by the ORA were corneal-compensated IOP (IOPcc), Goldmann-correlated IOP (IOPg), corneal hysteresis (CH), and corneal resistance factor (CRF). A difference more than 1.5 mmHg for IOP was considered as clinically relevant. Mean age of the patients was 12 years. Mean IOP (±standard deviation, SD) was 15.3 ± 2.8 mmHg (GAT), 15.5 ± 6.0 (NCT), 19.2 ± 7.0 (IOPg), and 21.1 ± 7.9 (IOPcc); (P = 0.001). Except for NCT vs. GAT (P = 1.0), the average IOP difference between each pair of measurements was clinically relevant. The 95% limits of agreements were - 10.2 to 10.3 mmHg (NCT vs. GAT), -7.8 to 15.3 (IOPg vs. GAT), and - 8.1 to 19.0 (IOPcc vs. GAT). The differences in IOP measurements increased significantly with higher average IOP values (r = 0.715, P = 0.001, for NCT vs. GAT; r = 0.802, P < 0.001, for IOPg vs. GAT; and r = 0.806, P < 0.001, for IOPcc vs. GAT). CH showed a significant association with differences in IOP measurements only for IOPcc vs. GAT (r = 0.830, P < 0.001). Mean IOP obtained by NCT was not significantly different from that of GAT, but ORA measured IOPs were significantly higher than both other devices.Keywords
This publication has 25 references indexed in Scilit:
- Corneal Thickness- and Age-Related Biomechanical Properties of the Cornea Measured with the Ocular Response AnalyzerInvestigative Ophthalmology & Visual Science, 2006
- Corneal Hysteresis and Intraocular Pressure Measurement in Children Using the Reichert Ocular Response AnalyzerAmerican Journal of Ophthalmology, 2006
- Evaluation of the Influence of Corneal Biomechanical Properties on Intraocular Pressure Measurements Using the Ocular Response AnalyzerJournal of Glaucoma, 2006
- Assessment of the accuracy and reliability of the Topcon CT80 non‐contact tonometerClinical and Experimental Optometry, 2006
- Influence of corneal biomechanical properties on intraocular pressure measurementJournal of Cataract & Refractive Surgery, 2005
- DIAGNOSIS AND MANAGEMENT OF PEDIATRIC GLAUCOMAOphthalmology Clinics of North America, 2001
- Human Corneal Thickness and Its Impact on Intraocular Pressure MeasuresSurvey of Ophthalmology, 2000
- Intraocular pressure: New perspectivesSurvey of Ophthalmology, 1990
- The non-contact tonometer. Its value and limitationsSurvey of Ophthalmology, 1980
- Long-term functional results in infantile glaucoma.1967