Assessment of an Automated Chemiluminescence Nitrogen Analyzer for Routine Use in Clinical Nutrition

Authors:
  • MWMalcolm F.E. West
    Malcolm F.E. West
  • AAAldo B.C. Acuti
    Aldo B.C. Acuti
  • RRRoger G. Rees
    Roger G. Rees
  • MHManjit K. Hunjan
    Manjit K. Hunjan
  • JWJoan D. Webster
    Joan D. Webster
  • PFPeter G. Frost
    Peter G. Frost
  • DSDavid B.A. Silk
    David B.A. Silk

Abstract
An automated method of chemiluminescence analysis of nitrogen used routinely for 4 yr. Liquid samples (urine, enteral, and parenteral feeds) required simple dilution, whereas feces required a modified acid‐digestion procedure, before analysis.For urine samples, the coefficient of variation was within batch from 0.9–3.6%, and between batch 4.3–7.6%. At a sample injection rate of 2 μl/sec, the useful dynamic range, for urine diluted 1:200, was 0–14 g N/liter. Precision for fecal nitrogen analysis was 3.8–6.7% for samples of low to high nitrogen content. The correlation between this technique and an established Kjeldahl method for fecal analysis was studied (r = 0.96, slope = 1.30). The discrepancy between the methods was due to inefficient conversion of nitrogen to NH4 + during Kjeldahl digestion of feces, rather than systematic errors in chemiluminescence analysis.Reliability was as good as for other automated clinical analyzers and sample cost was ca. £0.22. It has proved possible to analyze approximately 80 samples in the working day.The efficiency of measuring 24‐hr urine urea‐nitrogen (UUN) and total urine nitrogen (TUN) in patients on general wards was measured. Results were obtained on 87% of TPN days, but large variations were noted in UUN/TUN from 90% (average 75.7%) in patients receiving TPN, and from (Journal of Parenteral and Enteral Nutrition 12:100–106, 1988)

This publication has 21 references indexed in Scilit: