Potential dosing errors using portable prothrombin time monitoring devices
Authors:
- Daniel KurnikDKDaniel Kurnik
- Aharon LubetskyALAharon Lubetsky
- David EzraDEDavid Ezra
- Hillel HalkinHHHillel Halkin
- 1 July 2003
- journal article
- research article
- Published by Wolters Kluwer Health in Blood Coagulation & Fibrinolysis
- Vol. 14 (5) , 479-483
Abstract
Portable prothrombin time (PT) monitors facilitate the control of warfarin therapy. Few studies have compared the influence of using different monitors on dosage decisions. We determined the comparability of data generated by two portable PT monitors, Coaguchek S, (Roche Diagnostics Boehringer-Mannheim) and Hemochron Jr (International Technidyne Corporation Ltd.), with that of a reference laboratory. Simultaneous International Normalized Ratio (INR) measurements (portable monitor and laboratory) were performed in 193 consecutive patients receiving warfarin for at least 3 months. Agreement of measurements was assessed by both regression analysis and influence on dosage decisions in accordance with pre-defined criteria. The Coaguchek S versus laboratory INR regression line (n = 111; r2 = 0.88; P < 0.001) was close to the line of identity, while that of the Hemochron Jr (n = 82; r2 = 0.61; P < 0.001) was not. The overall proportion of dual INR measurements that fulfilled the clinical criteria of agreement was 90% for the Coaguchek S compared with 62% for the Hemochron Jr (P < 0.0001). For laboratory INRs 2.0-2.5, 2.6-4.0 and > 4.0, the proportions of portable measurements that satisfied the clinical criteria for the Coaguchek S versus the Hemochron Jr were 96 versus 63% (P < 0.001), 81 versus 45% (P < 0.04), and 67 versus 17% (P < 0.85), respectively. Warfarin dosing based solely on the portable devices would have resulted in unjustified dose increments in 22% of the patients with the Hemochron Jr device compared with 8% with the Coaguchek S monitor (chi2 = 4.43; P = 0.035). The Coaguchek S monitor provides measurements for INR values within the therapeutic range that agree well with the standard laboratory. The Hemochron Jr measurements result in different dosage adjustments even within the therapeutic range, but especially for INR values > 4.0. For both monitors, agreement of INR measurements with the standard decreases with increasing INR values.Keywords
This publication has 13 references indexed in Scilit:
- European Concerted Action on Anticoagulation (ECAA): International Normalized Ratio Variability of CoaguChek and TAS Point-of-Care Testing Whole Blood Prothrombin Time MonitorsThrombosis and Haemostasis, 2002
- A Prospective Controlled Trial Comparing Weekly Self-testing and Self-dosing with the Standard Management of Patients on Stable Oral AnticoagulationThrombosis and Haemostasis, 2000
- A Comparison of Point-of-Care Instruments Designed for Monitoring Oral Anticoagulation with Standard Laboratory MethodsThrombosis and Haemostasis, 2000
- A Comparison of INRs Determined with a Whole Blood Prothrombin Time Device and two International Reference Preparations for ThromboplastinThrombosis and Haemostasis, 2000
- A study of the variability seen in the international normalized ratio obtained using different sensitivity thromboplastin reagents on different instrument typesBlood Coagulation & Fibrinolysis, 1999
- Accuracy of a Portable International Normalization Ratio Monitor in Outpatients Receiving Long-Term Oral Anticoagulant TherapyThrombosis Research, 1998
- Evaluation of a portable prothrombin time monitor for home use by patients who require long-term oral anticoagulant therapyArchives of internal medicine (1960), 1993
- Are Capillary Whole Blood Coagulation Monitors Suitable for the Control of Oral Anticoagulant Treatment by the International Normalized Ratio?Thrombosis and Haemostasis, 1993
- Accuracy and precision of a portable anticoagulation monitor in a clinical settingArchives of internal medicine (1960), 1992
- Patient self-management of oral anticoagulation guided by capillary (fingerstick) whole blood prothrombin timesArchives of internal medicine (1960), 1989