Note: Multi-point measurement of |B| in the gas-dynamic trap with a spectral motional Stark effect diagnostic
Authors:
- 1 August 2011
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 82 (8) , 086105
Abstract
An upgraded spectral motional Stark effect diagnostic has been installed on the gas-dynamic trap (GDT) experiment to enable spatially resolved measurement of |B|. A new low-noise charge-coupled device detector, combined with enhancements of the diagnostic neutral beam, allows single-shot profile measurements. Previously only single-point motional Stark effect measurements were possible, and detector noise severely limited measurement precision, requiring multi-shot averaging. The plasma pressure profile in GDT is derived from the measured diamagnetic modification of |B| and used to examine the conditions of stable plasma confinement at high plasma pressure.Keywords
This publication has 5 references indexed in Scilit:
- Results of Recent Experiments on GDT Device After Upgrade of Heating Neutral BeamsFusion Science and Technology, 2010
- Advances in neutral-beam-based diagnostics on the Madison Symmetric Torus reversed-field pinch (invited)Review of Scientific Instruments, 2006
- Neutral beam injector for active plasma spectroscopyReview of Scientific Instruments, 2004
- Gas dynamic trap as high power 14 MeV neutron sourceFusion Engineering and Design, 2004
- Measurements of the radial profile of magnetic field in the Gas-Dynamic Trap using a motional Stark effect diagnosticReview of Scientific Instruments, 2003