Measuring airway dimensions during bronchoscopy using anatomical optical coherence tomography

Authors:

Abstract
Airway dimensions are difficult to quantify bronchoscopically because of optical distortion and a limited ability to gauge depth. Anatomical optical coherence tomography (aOCT), a novel imaging technique, may overcome these limitations. This study evaluated the accuracy ofaOCT against existing techniques in phantom, excised pig andin vivohuman airways.Three comparative studies were performed: 1) micrometer-derived area measurements in 10 plastic tubes were compared withaOCT-derived area; 2)aOCT-derived airway compliance curves from excised pig airways were compared with curves derived using an endoscopic technique; and 3) airway dimensions from the trachea to subsegmental bronchi were measured usingaOCT in four anaesthetised patients during bronchoscopy and compared with computed tomography (CT) measurements.Measurements in plastic tubes revealedaOCT to be accurate and reliable. In pig airways,aOCT-derived compliance measurements compared closely with endoscopic data. In human airways, dimensions measured withaOCT and CT correlated closely. Bland–Altman plots showed thataOCT diameter and area measurements were higher than CT measurements by 7.6% and 15.1%, respectively.Airway measurements usingaOCT are accurate, reliable and compare favourably with existing imaging techniques. UsingaOCT with conventional bronchoscopy allows real-time measurement of airway dimensions and could be useful clinically in settings where knowledge of airway calibre is required.