Porous titanium alloy for fixation of knee prostheses
Authors:
- 1 January 1974
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research
- Vol. 8 (4) , 319-328
Abstract
A porous titanium alloy called void metal composite (VMC) was investigated for use in knee prosthesis fixation. The interfacial ingrowth fixation of VMC intrameduallary, subchondral, and variable density implants in goat femora was evaluated biomechanically and histologically at six weeks and five months. Biomechanical tests were done under static shear and dynamic, cyclic (fatigue) loading conditions. Readily identifiable bone, strong fixation, and excellent biocompatibility were typical. Some subarticular implants appeared to induce normal hyaline cartilage overgrowth.Keywords
This publication has 1 reference indexed in Scilit:
- Biomechanical investigations of bone-porous carbon and porous metal interfacesJournal of Biomedical Materials Research, 1973