Abstract
There is growing concern about the accuracy of trace elemental analysis of ambient particulate matter (PM) samples. This has become important because ambient PM concentrations have decreased over the years, and the lower filter loadings result in difficulties in accurate analysis. The performance of energy-dispersive X-ray reflectance spectrometry was evaluated at Harvard School of Public Health using several methodologies, including intercomparison between two other laboratories. In reanalysis of standard films as unknown samples following calibration, the HSPH ED XRF measurements represented good performance: 2% errors in precision and 4% errors in accuracy. Replicate analysis of ambient air filters with low PM2.5 levels indicated that S, K, Fe, and Ca showed excellent reproducibility, most other quantifiable elements were below 15% error, and the elements with larger percent of flagged measurements had less in precision. Results from the interlaboratory comparison demonstrated that most quantifiable elements, except Na and Al, were quite comparable for the three laboratories. Na performance could be validated from the stoichiometry of Na to Cl of indoor PM2.5 filter samples.Keywords
Funding Information
- U.S. Environmental Protection Agency (RD-83479801)
This publication has 20 references indexed in Scilit:
- Reanalysis of Archived IMPROVE PM2.5 Samples Previously Analyzed over a 15-Year PeriodEnvironmental Science & Technology, 2012
- Toxicological evaluation of realistic emission source aerosols (TERESA): summary and conclusionsInhalation Toxicology, 2011
- Toxicological Evaluation of Realistic Emission Source Aerosols (TERESA)—Power plant studies: assessment of breathing patternInhalation Toxicology, 2011
- Atmospheric trace element concentrations in total suspended particles near Paris, FranceAtmospheric Environment, 2010
- Comparison of ED-XRF and LA-ICP-MS with the European reference method of acid digestion-ICP-MS for the measurement of metals in ambient particulate matterAccreditation and Quality Assurance, 2010
- Atomic spectrometry update. X-ray fluorescence spectrometryJournal of Analytical Atomic Spectrometry, 2007
- COMPARATIVE PULMONARY TOXICITY OF VARIOUS SOLUBLE METALS FOUND IN URBAN PARTICULATE DUSTSExperimental Lung Research, 2002
- X-ray fluorescence analysis of silicate rocks using fused glass discs and a side-window Rh source tube: accuracy, precision and reproducibilityChemical Geology, 1993
- Development and evaluation of a glass honeycomb denuder/filter pack system to collect atmospheric gases and particlesEnvironmental Science & Technology, 1993