Use Caution When Applying Magnets to Pacemakers or Defibrillators for Surgery
Authors:
- Marc A. RoznerMRMarc A. Rozner
- 1 August 2013
- journal article
- case report
- Published by Wolters Kluwer Health in Anesthesia & Analgesia
- Vol. 117 (2) , 422-427
Abstract
The application of a magnet to a pacemaker (intended to cause asynchronous pacing) or implanted cardioverter defibrillator (intended to prevent shocks) during surgery without a clear understanding of actual magnet function(s) or precautions can have unexpected, untoward, or harmful consequences. In this report, we present 3 cases in which inadequate assessment of cardiac implanted electronic device (CIED) function, coupled with magnet application, contributed to or resulted in inappropriate antitachycardia pacing or shocks, CIED damage, or patient injury. Although these cases might be rare, they reinforce the need for a timely, detailed preoperative review of CIED function and programming as recommended by the American Society of Anesthesiologists and the Heart Rhythm Society.Keywords
This publication has 16 references indexed in Scilit:
- Reduction in Inappropriate Therapy and Mortality through ICD ProgrammingNew England Journal of Medicine, 2012
- Implantable cardioverter defibrillators: risks accompany the life-saving benefitsHeart, 2012
- Society position statementCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 2012
- Clinical applications of magnets on cardiac rhythm management devicesEP Europace, 2011
- Inappropriate Implantable Cardioverter-Defibrillator ShocksJACC, 2011
- Current Perioperative Management of the Patient With a Cardiac Rhythm Management DeviceSeminars in Cardiothoracic and Vascular Anesthesia, 2009
- Prognostic Importance of Defibrillator Shocks in Patients with Heart FailureNew England Journal of Medicine, 2008
- Pacemakers and implantable cardioverter defibrillatorsAnaesthesia, 2006
- Unintentional Deactivation of Implantable Cardioverter-Defibrillators in Health Care SettingsMayo Clinic Proceedings, 2002
- Analysis of Troponin I Levels After Spontaneous Implantable Cardioverter Defibrillator ShocksJournal of Cardiovascular Electrophysiology, 2002