Quasi-static modeling of the da Vinci instrument
Authors:
- 1 September 2014
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 21530858,p. 1308-1313
Abstract
Two simplified quasi-static models for the da Vinci instrument are proposed which take into account distributed frictions and compliance of the tendons. These models are derived from static analysis of the interaction of the tendons with a curved surface. The key parameters of the models are identified, and the performance of the models is experimentally evaluated. Experimental results obtained suggest that a weighted combination of the outputs of the two models provides a sufficiently close estimate of the output torque of the da Vinci instrument.Keywords
This publication has 12 references indexed in Scilit:
- The hand of the DLR Hand Arm System: Designed for interactionThe International Journal of Robotics Research, 2012
- Modeling, Identification, and Control of Tendon-Based Actuation SystemsIEEE Transactions on Robotics, 2011
- Modeling of Transmission Characteristics Across a Cable-Conduit SystemIEEE Transactions on Robotics, 2010
- Design of tendon-driven robotic fingers: Modeling and control issues2010
- Control of cable actuated devices using smooth backlash inverse2010
- Force sensing and its application in minimally invasive surgery and therapy: A surveyProceedings of the Institution of Mechanical Engineers, Part C: Journal of Mechanical Engineering Science, 2010
- Haptic feedback in robot-assisted minimally invasive surgeryCurrent Opinion in Urology, 2009
- Design of an endoluminal NOTES robotic system2007
- An investigation of the transmission system of a tendon driven robot hand2002
- The Intuitive/sup TM/ telesurgery system: overview and application2002