The performance properties of miniature piezoelectric polymer ultrasonic hydrophone probes are described, their applications as reference hydrophones and for in vivo ultrasonic measurement of tissue properties are discussed, and the advantages of a needle-like construction are pointed out. A novel calibration technique, based on Time Delay Spectrometry, allows the hydrophone probe free field parameters to be determined as a continuous function of frequency in reverberant environments, while maintaining outstanding signal-to-noise ratio.Keywords
This publication has 18 references indexed in Scilit:
- Calibration of miniature ultrasonic receivers using a planar scanning techniqueThe Journal of the Acoustical Society of America, 1982
- Angular response of miniature ultrasonic hydrophonesMedical Physics, 1982
- Field patterns of pulsed, focused, ultrasonic radiators in attenuating and nonattenuating mediaThe Journal of the Acoustical Society of America, 1982
- A comparison of two calibration methods for ultrasonic hydrophonesUltrasound in Medicine & Biology, 1982
- Simplified technique for the calibration and use of a miniature hydrophone in intensity measurements of pulsed ultrasound fieldsThe Journal of the Acoustical Society of America, 1981
- Piezoelectric polymer probe for ultrasonic applicationsThe Journal of the Acoustical Society of America, 1981
- Tissue characterizationUltrasound in Medicine & Biology, 1981
- Effective area to be used in diffraction correctionsThe Journal of the Acoustical Society of America, 1980
- Assessment of errors in intensity measurements of pulse echo ultrasound using miniature hydrophonesMedical Physics, 1979
- Piezoelectric polymer receiving arrays for ultrasonic applicationsThe Journal of the Acoustical Society of America, 1977