Evaluation of a Minimally Invasive System for Measuring Glucose Area under the Curve during Oral Glucose Tolerance Tests: Usefulness of Sweat Monitoring for Precise Measurement
Open Access
- 1 May 2013
- journal article
- research article
- Published by SAGE Publications in Journal of Diabetes Science and Technology
- Vol. 7 (3) , 678-688
Abstract
We developed a system for measuring glucose area under the curve (AUC) using minimally invasive interstitial fluid extraction technology (MIET). Sweat contamination during interstitial fluid glucose (IG) extraction affects the accuracy of glucose AUC measurement, because this technology uses extracted sodium ion levels as an internal standard. Therefore, we developed a sweat monitoring patch to reduce this effect and investigated its efficacy in volunteers undergoing oral glucose tolerance tests (OGTTs). Materials and Methods: Fifty diabetes mellitus inpatients and 10 healthy subjects undergoing the 75 g OGTT were included. Two sites on the forearm were pretreated with microneedle arrays, then hydrogels for interstitial fluid extraction were placed on the treated sites. Simultaneously, hydrogels for sweat monitoring were placed on untreated sites near the treated sites. Plasma glucose (PG) levels were measured every 30 min for 2 h to calculate reference AUC values. Using MIET, IG AUC was calculated from extracted glucose and sodium ion levels after attachment of the hydrogel for 2 h. Good correlation between IG AUC measurements using MIET and reference AUCs measured using PG levels was confirmed over a wide AUC range (202–610 mg/h/dl) after correction for the sweat-induced error detected by the hydrogel patches on the nonpretreated skin. Strong correlation between IG AUC and peak glucose levels indicates that glucose spikes can be easily detected by this system. We confirmed the effectiveness of a sweat monitoring patch for precise AUC measurement using MIET. This novel, easy-to-use system has potential for glucose excursion evaluation in daily clinical practice.Keywords
This publication has 22 references indexed in Scilit:
- A Minimally Invasive System for Glucose Area Under the Curve Measurement Using Interstitial Fluid Extraction Technology: Evaluation of the Accuracy and Usefulness with Oral Glucose Tolerance Tests in Subjects With and Without DiabetesDiabetes Technology & Therapeutics, 2012
- Management of Hyperglycemia in Type 2 Diabetes: A Patient-Centered ApproachDiabetes Care, 2012
- Correlation Between Sweat Glucose and Blood Glucose in Subjects with DiabetesDiabetes Technology & Therapeutics, 2012
- International clinical harmonization of glycated hemoglobin in Japan: From Japan Diabetes Society to National Glycohemoglobin Standardization Program valuesJournal of Diabetes Investigation, 2012
- Standards of Medical Care in Diabetes—2012Diabetes Care, 2011
- Measurement of Glucose Area Under the Curve Using Minimally Invasive Interstitial Fluid Extraction Technology: Evaluation of Glucose Monitoring Concepts Without Blood SamplingDiabetes Technology & Therapeutics, 2011
- Dutogliptin, a selective DPP4 inhibitor, improves glycaemic control in patients with type 2 diabetes: a 12‐week, double‐blind, randomized, placebo‐controlled, multicentre trialDiabetes, Obesity and Metabolism, 2010
- Continuous Glucose Monitoring Health OutcomesDiabetes Technology & Therapeutics, 2009
- Hyperglycaemia and mortality from all causes and from cardiovascular disease in five populations of Asian originDiabetologia, 2004
- Is the Current Definition for Diabetes Relevant to Mortality Risk From All Causes and Cardiovascular and Noncardiovascular Diseases?Diabetes Care, 2003