Preserved Low-Frequency Hearing Following 20-mm Cochlear Implantation

Authors:
  • MMMyles F. Melton
    Myles F. Melton
  • HSHannah Shonfield
    Hannah Shonfield
  • MKMichelle Kraskin
    Michelle Kraskin
  • JWJennifer Wolf
    Jennifer Wolf

Abstract
To determine the preservation of pure tone thresholds for a series of cochlear implant patients who underwent atraumatic round window insertion with a new thin 22-mm electrode to a 20-mm depth.Retrospective, within-subject repeated measures design.Tertiary care hospital.Nine sequential cochlear implant patients with functional preoperative acoustic hearing (defined as low-frequency pure tone average thresholds ≤90 dB).Therapeutic.Pure tone thresholds after cochlear implantation.Low-frequency hearing preservation was achieved in all cases. Less than 10 dB average change was seen for low-frequency pure tone average measures (125, 250, 500, and 1,000 Hz) at 6 and 12 months following cochlear implantation.Atraumatic round window insertion with a thin 22-mm electrode to a 20-mm depth results in dramatically high levels of acoustic hearing preservation among CI patients.