Abstract
We report on the use of a kinetic-inductance traveling-wave parametric amplifier (KITWPA) as the first amplifier in the readout chain of a microwave superconducting quantum interference device (SQUID) multiplexer (μmux). This μmux is designed to multiplex signals from arrays of low temperature detectors, such as superconducting transition-edge sensor microcalorimeters. When modulated with a periodic flux-ramp to linearize the SQUID response, the flux noise improves, on average, from 1.6 μΦ0/Hz with the KITWPA off to 0.77 μΦ0/Hz with the KITWPA on. When statically biasing the μmux to the maximally flux-sensitive point, the flux noise drops from 0.45 μΦ0/Hz to 0.2 μΦ0/Hz. We validate this new readout scheme by coupling a transition-edge sensor microcalorimeter to the μmux and detecting background radiation. The combination of μmux and KITWPA provides a variety of new capabilities including improved detector sensitivity and more efficient bandwidth utilization.
All Related Versions
Funding Information
  • National Institute of Standards and Technology (NIST Program on Scalable Superconducting Computing)
  • National Aeronautics and Space Administration (NNH18ZDA001N-APRA)
  • U.S. Department of Energy (89243020SSC000058)

This publication has 19 references indexed in Scilit: