Imaging System for Intravascular Ultrasonography Using Pulse Compression Technique
Authors:
- Zhiqiang HuZHZhiqiang Hu
- Tadashi MoriyaTMTadashi Moriya
- Yoshikatsu TanahashiYTYoshikatsu Tanahashi
- 1 May 2001
- journal article
- Published by IOP Publishing in Japanese Journal of Applied Physics
- Vol. 40 (5S)
Abstract
We have developed an imaging system that applies FM-chirp pulse compression technology for use in intraductal ultrasonography, or intravascular ultrasonography. The pulse compression technique attempts to raise resolving power by using average power effectively under peak power restriction. This paper describes a method that use pulse compression technology in intravascular ultrasonography. In this paper, the following are reported: 1) A method to use the FM-chirp pulse compression technique in an imaging system is proposed. 2) A transmission line suited for intravascular ultrasonography is constructed using the L(0, 1) mode of the Pochhmmer-Chree wave propagating in a fused quartz rod coated with carbon. 3) A receiving signal from a target is compressed to give a shot pulse.Keywords
This publication has 3 references indexed in Scilit:
- A Method to Separate the Transmitting Signal and the Receiving Signal in FM-Chirp Pulse Compression SystemsJapanese Journal of Applied Physics, 2000
- A 20 MHz single-element ultrasonic probe using 0.91Pb(Zn/sub 1/3/Nb/sub 2/3/)O/sub 3/-0.09PbTiO/sub 3/ single crystalIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1998
- Wire-Type Dispersive Ultrasonic Delay LinesIRE Transactions on Ultrasonic Engineering, 1960