Imaging System for Intravascular Ultrasonography Using Pulse Compression Technique

Authors:
  • ZHZhiqiang Hu
    Zhiqiang Hu
  • TMTadashi Moriya
    Tadashi Moriya
  • YTYoshikatsu Tanahashi
    Yoshikatsu Tanahashi

Abstract
We have developed an imaging system that applies FM-chirp pulse compression technology for use in intraductal ultrasonography, or intravascular ultrasonography. The pulse compression technique attempts to raise resolving power by using average power effectively under peak power restriction. This paper describes a method that use pulse compression technology in intravascular ultrasonography. In this paper, the following are reported: 1) A method to use the FM-chirp pulse compression technique in an imaging system is proposed. 2) A transmission line suited for intravascular ultrasonography is constructed using the L(0, 1) mode of the Pochhmmer-Chree wave propagating in a fused quartz rod coated with carbon. 3) A receiving signal from a target is compressed to give a shot pulse.

This publication has 3 references indexed in Scilit: