Experimental verification of IMPT treatment plans in an anthropomorphic phantom in the presence of delivery uncertainties
Authors:
- M CasiraghiMCM Casiraghi
- S LorentiniSLS Lorentini
- B RombiBRB Rombi
- A J LomaxALA J Lomax
- 27 June 2011
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 56 (14) , 4415-4431
Abstract
Clinically relevant intensity modulated proton therapy (IMPT) treatment plans were measured in a newly developed anthropomorphic phantom (i) to assess plan accuracy in the presence of high heterogeneity and (ii) to measure plan robustness in the case of treatment uncertainties (range and spatial). The new phantom consists of five different tissue substitute materials simulating different tissue types and was cut into sagittal planes so as to facilitate the verification of co-planar proton fields. GafChromic films were positioned in the different planes of the phantom, and 3D-IMPT and distal edge tracking (DET) plans were delivered to a volume simulating a skull base chordoma. In addition, treatments planned on CTs of the phantom with HU units modified were delivered to simulate systematic range uncertainties (range-error treatments). Finally, plans were delivered with the phantom rotated to simulate spatial errors. Results show excellent agreement between the calculated and the measured dose distribution: >99% and 98% of points with a gamma value <1 (3%/3 mm) for the 3D-IMPT and the DET plan, respectively. For both range and spatial errors, the 3D-IMPT plan was more robust than the DET plan. Both plans were more robust to range than to the spatial uncertainties. Finally, for range error treatments, measured distributions were compared to a model for predicting delivery errors in the treatment planning system. Good agreement has been found between the model and the measurements for both types of IMPT plan.Keywords
This publication has 35 references indexed in Scilit:
- Is it necessary to plan with safety margins for actively scanned proton therapy?Physics in Medicine & Biology, 2011
- The influence of the optimization starting conditions on the robustness of intensity-modulated proton therapy plansPhysics in Medicine & Biology, 2010
- LET dependence of GafChromic films and an ion chamber in low-energy proton dosimetryPhysics in Medicine & Biology, 2009
- Experiences at the Paul Scherrer Institute With a Remote Patient Positioning Procedure for High-Throughput Proton Radiation TherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008
- Performance of a fluorescent screen and CCD camera as a two-dimensional dosimetry system for dynamic treatment techniquesMedical Physics, 2000
- An application of GafChromic MD-55 film for 67.5 MeV clinical proton beam dosimetryPhysics in Medicine & Biology, 1999
- Calculation of the uncertainty in the dose delivered during radiation therapyMedical Physics, 1985
- Spot scanning system for proton radiotherapyMedical Physics, 1980
- Flattening of proton dose distributions for large‐field radiotherapyMedical Physics, 1977
- Range modulators for protons and heavy ionsNuclear Instruments and Methods, 1975