Calibration of Troitsk Nu-Mass Detector Readout Electronics by Signal Digital Filters
Authors:
- 15 May 2023
- journal article
- research article
- Published by The Russian Academy of Sciences in ЖУРНАЛ ЭКСПЕРИМЕНТАЛЬНОЙ И ТЕОРЕТИЧЕСКОЙ ФИЗИКИ
- Vol. 163 (5) , 641-645
Abstract
We present the results of tuning and calibration of the detector electronics in the signal digitization mode. The goal of the experiment is to search for a possible sterile neutrino signature in tritium beta-decay. The read-out electronics work in direct oscilloscope mode, which requires to optimize time frame the with the goal to minimize noise and energy resolution. We use a 7-pixel silicon drift detector (SDD) and a CMOS charge sensitive preamplifier with very low integration capacitor. Amplifier forms a slowly rising output shape and operates in pulse-reset mode. The 125 MHz ADC digitizes the signals. Using calibration data from Fe55 and Am241 gamma sources we check triangular and trapezoid digital filters to obtain the best noise and energy resolution performance. We are also examining the option to differentiate the output signal.Keywords
All Related Versions
This publication has 5 references indexed in Scilit:
- Measurements with a TRISTAN prototype detector system at the “Troitsk nu-mass” experiment in integral and differential modeJournal of Instrumentation, 2019
- Three stages of the NICA accelerator complexThe European Physical Journal A, 2016
- Optimized readout methods of silicon drift detectors for high-resolution X-ray spectroscopyNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2006
- Silicon drift detectors for high count rate X-ray spectroscopy at room temperatureNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2001
- Ballistic deficit correction in semiconductor detector spectrometersIEEE Transactions on Nuclear Science, 1988