A 3-D finite element analysis of the natural frequencies of vibration of a stapes prosthesis replacement reconstruction of the middle ear
Authors:
- 1 February 1995
- journal article
- Published by Wiley in Clinical Otolaryngology
- Vol. 20 (1) , 36-44
Abstract
In this work, the natural frequencies of vibration of two different stapes prosthesis replacement reconstructions have been analysed using the finite element method. Prosthesis 1 was constructed of fine stainless steel wire and a Teflon base, while prosthesis 2 was made entirely of Teflon. The results have indicated that generally, the first natural frequency of vibration falls as the prostheses become larger and more bulky. However, the fall in the first natural for prosthesis 1 was modest when compared with that of the isolated tympanic membrane. An important variable influencing frequencies and mode of vibration of the reconstruction was the tightness of fit of the prostheses within the remaining ossicular chain. The tightness of fit in this work was modelled as a range of spring constants applied at the stapes pseudo footplate, together with a series of joint elements at the incus tip/prosthesis head for the Teflon implant. When these spring constraints were below approximately 10 N/mm, the reconstruction exhibited normal mode 1 vibration behaviour, but at larger spring values, an abnormal mode 1 became established resembling the normal mode 2 vibration characteristics. The formation of new geometries and surfaces following reconstruction, introduces new constraints between umbo and stapes footplate particularly at the natural/artificial interfaces. These unnatural constraints may inhibit and modify the natural movements normally occurring at the stapes footplate leading to abnormal modes of vibration.Keywords
This publication has 11 references indexed in Scilit:
- Natural frequencies of vibration of a fibre supported human tympanic membrane analysed by the finite element methodClinical Otolaryngology, 1993
- Analysis of dynamic behavior of human middle ear using a finite-element methodThe Journal of the Acoustical Society of America, 1992
- A dynamic and natural frequency analysis of the Fisch II spandrel using the finite element methodClinical Otolaryngology, 1992
- Mechanics and materials in middle ear reconstructionClinical Otolaryngology, 1991
- A finite element analysis of the natural frequencies of vibration of the human tympanic membrane. Part IBritish Journal of Audiology, 1990
- The tympanic membrane in cross section: a finite element analysisThe Journal of Laryngology & Otology, 1988
- Present management of the reconstruction of the middle earClinical Otolaryngology, 1987
- On the undamped natural frequencies and mode shapes of a finite-element model of the cat eardrumThe Journal of the Acoustical Society of America, 1983
- Modeling of the cat eardrum as a thin shell using the finite-element methodThe Journal of the Acoustical Society of America, 1978
- Tympanic-Membrane Vibrations in Human Cadaver Ears Studied by Time-Averaged HolographyThe Journal of the Acoustical Society of America, 1972