Evaluation of the Nallasamy formula: a stacking ensemble machine learning method for refraction prediction in cataract surgery
Open Access
- 4 April 2022
- journal article
- research article
- Published by BMJ in British Journal of Ophthalmology
- Vol. 107 (8) , 1066-1071
Abstract
Aims To develop a new intraocular lens power selection method with improved accuracy for general cataract patients receiving Alcon SN60WF lenses. Methods and analysis A total of 5016 patients (6893 eyes) who underwent cataract surgery at University of Michigan’s Kellogg Eye Center and received the Alcon SN60WF lens were included in the study. A machine learning-based method was developed using a training dataset of 4013 patients (5890 eyes), and evaluated on a testing dataset of 1003 patients (1003 eyes). The performance of our method was compared with that of Barrett Universal II, Emmetropia Verifying Optical (EVO), Haigis, Hoffer Q, Holladay 1, PearlDGS and SRK/T. Results Mean absolute error (MAE) of the Nallasamy formula in the testing dataset was 0.312 Dioptres and the median absolute error (MedAE) was 0.242 D. Performance of existing methods were as follows: Barrett Universal II MAE=0.328 D, MedAE=0.256 D; EVO MAE=0.322 D, MedAE=0.251 D; Haigis MAE=0.363 D, MedAE=0.289 D; Hoffer Q MAE=0.404 D, MedAE=0.331 D; Holladay 1 MAE=0.371 D, MedAE=0.298 D; PearlDGS MAE=0.329 D, MedAE=0.258 D; SRK/T MAE=0.376 D, MedAE=0.300 D. The Nallasamy formula performed significantly better than seven existing methods based on the paired Wilcoxon test with Bonferroni correction (pConclusions The Nallasamy formula (available at https://lenscalc.com/) outperformed the seven other formulas studied on overall MAE, MedAE, and percentage of eyes within 0.5 D of prediction. Clinical significance may be primarily at the population level.Keywords
All Related Versions
Funding Information
- National Eye Institute, Bethesda, MD (1R01EY026641-01A1)
- Lighthouse Guild, New York, NY (None)
This publication has 30 references indexed in Scilit:
- Indications and Outcomes of Intraocular Lens Exchange During a Recent 5-Year PeriodAmerican Journal of Ophthalmology, 2014
- Errata in printed Hoffer Q formulaJournal of Cataract & Refractive Surgery, 2007
- Reply : Errata in printed Hoffer Q formulaJournal of Cataract & Refractive Surgery, 2007
- CorrectionJournal of Cataract & Refractive Surgery, 1994
- The Hoffer Q formula: A comparison of theoretic and regression formulasJournal of Cataract & Refractive Surgery, 1993
- Stacked generalizationNeural Networks, 1992
- ErratumJournal of Cataract & Refractive Surgery, 1990
- ECG Abnormalities and Systemic Diseases in Patients Having Cataract or Glaucoma SurgeryJournal of Cataract & Refractive Surgery, 1990
- Development of the SRK/T intraocular lens implant power calculation formulaJournal of Cataract & Refractive Surgery, 1990
- A three-part system for refining intraocular lens power calculationsJournal of Cataract & Refractive Surgery, 1988