Performance evaluation of a novel multi-pinhole collimator for dopamine transporter SPECT
- 14 August 2020
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 65 (16) , 165015
Abstract
There is a tradeoff between spatial resolution and count sensitivity in SPECT with conventional collimators. Multi-pinhole (MPH) collimator technology has potential for concurrent improvement of resolution and sensitivity in clinical SPECT of ‘small’ organs. This study evaluated a novel MPH collimator specifically designed for dopamine transporter (DAT) SPECT with a triple-head SPECT camera. Count sensitivity was measured with a 99m Tc point source placed on the lattice points of a 1 cm grid covering the whole field-of-view (FOV). Spatial resolution was assessed with a Derenzo type hot rod phantom. An anthropomorphic striatum phantom was scanned with total activity representative of a typical patient scan and different striatum-to-background activity concentration ratios. Recovery of striatum-to-background contrast was assessed by the contrast-recovery-coefficient. Measurements were repeated with double-head SPECT with fan-beam or low-energy-high-resolution-high-sensitivity (LEHRHS) collimators. A patient referred to DAT SPECT because of suspicion of Parkinson’s disease was scanned with both LEHRHS and MPH collimators after a single tracer injection. The axial MPH sensitivity profile was approximately symmetrical around its peak, although it was shifted 7 cm towards the patient to simplify positioning. Peak sensitivity of the triple-head MPH system in the center of the FOV was 620 cps MBq −1 compared to 225 cps MBq −1 for the double-head fan-beam system. Sensitivity of the MPH system decreased towards the edges of the FOV. The full width of the sensitivity profile at 200 cps MBq −1 was 21 cm transaxially and 11 cm axially. In MPH SPECT of the Derenzo phantom all rods with ≥ 5 mm diameter were clearly visible. MPH SPECT improved striatal contrast recovery by ≥ 20% compared to fan-beam SPECT. The patient scan demonstrated good image quality of MPH SPECT with almost PET-like delineation of putamen and caudate nucleus. SPECT with dedicated MPH collimators provides considerable improvement of the resolution-sensitivity tradeoff in DAT SPECT compared to SPECT with fan-beam or LEHRHS collimators.Keywords
This publication has 64 references indexed in Scilit:
- Advances in Preclinical SPECT InstrumentationJournal of Nuclear Medicine, 2012
- SNM Practice Guideline for Dopamine Transporter Imaging with 123I-Ioflupane SPECT 1.0Journal of Nuclear Medicine, 2011
- Small animal imaging with multi-pinhole SPECTMethods, 2009
- Multi-pinhole collimator design for small-object imaging with SiliSPECT: a high-resolution SPECTPhysics in Medicine & Biology, 2008
- Design and performance of a multi-pinhole collimation device for small animal imaging with clinical SPECT and SPECT–CT scannersPhysics in Medicine & Biology, 2008
- Quantitative [123I]FP-CIT pinhole SPECT imaging predicts striatal dopamine levels, but not number of nigral neurons in different mouse models of Parkinson's diseaseNeuroImage, 2007
- Movies of dopamine transporter occupancy with ultra-high resolution focusing pinhole SPECTMolecular Psychiatry, 2007
- Pinhole SPECT imaging of dopamine transporters correlates with dopamine transporter immunohistochemical analysis in the MPTP mouse model of Parkinson's diseaseNeuroImage, 2005
- AGEING AND PARKINSON'S DISEASE: SUBSTANTIA NIGRA REGIONAL SELECTIVITYBrain, 1991
- Bayesian reconstructions from emission tomography data using a modified EM algorithmIEEE Transactions on Medical Imaging, 1990