Dependence of scatter on atomic number for x rays from tungsten and molybdenum anodes in the mammographic energy range
Authors:
- R. J. AusRAR. J. Aus
- L. A. DeWerdLDL. A. DeWerd
- D. W. PearsonDPD. W. Pearson
- J. A. MickaJMJ. A. Micka
- K. H. NgKNK. H. Ng
- 12 July 1999
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 26 (7) , 1306-1311
Abstract
A study was done to determine the relative amounts of scatter for the following materials with atomic numbers ranging from Z=6 to Z=82: C, Al, Ti, Fe, Cu, Zn, Zr, Y, Mo, Ta, and Pb. Measurements were performed for each material on two constant potential x-ray units--one fitted with a molybdenum (Mo) anode-Mo filter and the other with a tungsten (W) anode-aluminum (Al) filter (medium filtration) at 30 kVp. Theoretical calculations were also performed for each anode to explain the scatter behavior and to aid in predicting the behavior for materials where measurements were not made. There was good agreement between the theoretical calculations and the experimental data.Keywords
This publication has 5 references indexed in Scilit:
- Comparison of exposure standards in the mammography x‐ray regionMedical Physics, 1997
- Mammography exposure standard: Design and characterization of free-air ionization chamberReview of Scientific Instruments, 1995
- Introduction to Radiological Physics and Radiation Dosimetry1986
- Measurements of diagnostic x‐ray backscatter by a novel ion chamber methodMedical Physics, 1982
- Backscatter factors in the mammographic energy rangeMedical Physics, 1977