An Alternate Method of Diluting Dissolved Organic Carbon Seawater Samples for 14C Analysis
Open Access
- 1 January 2010
- journal article
- research article
- Published by Cambridge University Press (CUP) in Radiocarbon
- Vol. 52 (3) , 1224-1229
Abstract
We present a time-saving modification to the ultraviolet (UV) oxidation method for analyzing dissolved organic carbon (DOC) concentration, Δ14C, and δ13C measurements in seawater and standard materials. A low background (∼0.2 ± 0.2 μM) was reported for pre-irradiated Milli-Q (MQ) water that was used to dilute samples for DOC 14C analysis (Beaupré et al. 2007). We use MQ water without pre-irradiation (background ∼0.9 ± 0.2 μM) to dilute the sample. This method is suitable for small-volume, high-concentration samples (mass of sample DOC overwhelms mass of MQ water DOC). An acceptable precision of Δ14C measurements (5–9‰) is maintained. This revised method reduces the preparation time for diluted DOC Δ14C samples from 2 days to 1 day.Keywords