Intercomparison of Calibration Systems for AC Shunts Up to Audio Frequencies
Authors:
- T. TsuchiyamaTTT. Tsuchiyama
- T. TadokoroTTT. Tadokoro
- B.C. WaltripBWB.C. Waltrip
- T.L. NelsonTNT.L. Nelson
- 21 March 2005
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Instrumentation and Measurement
- Vol. 54 (2) , 507-511
Abstract
An intercomparison of calibration systems for AC shunts up to audio frequencies (10 kHz) between the National Research Council of Canada, Japan Electric Meters Inspection Corporation, and the National Institute of Standards and Technology, is presented. The comparison was implemented with a transfer standard of 10 A, 0.1 /spl Omega/ calculable AC/DC shunt, designed by Japan Electric Meters Inspection Corporation. The results indicate that there are no significant differences in the overall accuracy of calibration systems for AC shunts at frequencies up to 10 kHz in each laboratory.Keywords
This publication has 7 references indexed in Scilit:
- A four terminal current shunt with calculable AC response2004
- Development of a high precision AC standard shunt for AC power measurement2003
- NRC-NIST intercomparison of calibration systems for current transducers with a voltage output at power frequenciesIEEE Transactions on Instrumentation and Measurement, 2003
- A current comparator bridge for power measurementIEEE Transactions on Instrumentation and Measurement, 1976
- Measurement of short circuit load losses in large three phase power transformers using an alternating current comparator bridgeIEEE Transactions on Power Apparatus and Systems, 1975
- The Development and Performance of Current Comparators for Audio FrequenciesIEEE Transactions on Instrumentation and Measurement, 1965
- A design for resistors of calculable a.c./d.c. resistance ratioProceedings of the Institution of Electrical Engineers, 1963