Double Peak Axial Length Measurement Signal in Cataract Patients with Epiretinal Membrane

Authors:

Abstract
Purpose: To evaluate the accuracy of axial length (AL) measurement for intraocular lens (IOL) calculation in patients with cataract and epiretinal membrane (ERM).Methods: This prospective, cross-sectional study was performed in cataract patients with ERM. All subjects were sent for standard optical biometry, prepared for cataract surgery. Signals of AL measurement were detected as double peaks and recorded as AL1 (first peak), and AL2 (second peak). The IOL power was calculated from AL1 and AL2, and reported as IOL1 and IOL2. The IOL2 was chosen for cataract surgery in all cases. Postoperative predictive errors were compared between IOL1 and IOL2. Results: Thirty-seven eyes from 37 patients were included. Mean AL1 was significantly shorter than AL2 (23.13±1.28 vs. 23.60±1.34 mm, pConclusions: AL measurement in ERM can be detected as a double peak signal during biometric measurement. The IOL power calculated from the first and second peak signals are significantly different. However, the IOL power derived from the second peak signal provides better refractive outcomes. The results suggest that the second peak signal represent an accurate AL measurement.