THE MEASUREMENT OF AIRBORNE GLUTARALDEHYDE BY HIGH-PERFORMANCE LIQUID CHROMATOGRAPHY

Authors:

Abstract
A published method for measuring personal exposure to glutaraldehyde has been evaluated. A change to the composition of the eluent used in the high-performance liquid chromatography analysis is suggested along with a more robust calibration procedure. Test atmospheres of glutaraldehyde were used to show that sampling efficiency is unaffected by humidity changes in the range 20–70% using sampling flow rates of up to 500 ml min−1. Samples can be stored at room temperature for up to 4 weeks after collection The method has a detection limit of about 3 ppb for a 21 sample equivalent to about 1/65th of the current occupational exposure limit for glutaraldehyde