SAR simulations for high‐field MRI: How much detail, effort, and accuracy is needed?
Open Access
- 18 May 2012
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 69 (4) , 1157-1168
Abstract
Accurate prediction of specific absorption rate (SAR) for high field MRI is necessary to best exploit its potential and guarantee safe operation. To reduce the effort (time, complexity) of SAR simulations while maintaining robust results, the minimum requirements for the creation (segmentation, labeling) of human models and methods to reduce the time for SAR calculations for 7 Tesla MR‐imaging are evaluated. The geometric extent of the model required for realistic head‐simulations and the number of tissue types sufficient to form a reliable but simplified model of the human body are studied. Two models (male and female) of the virtual family are analyzed. Additionally, their position within the head‐coil is taken into account. Furthermore, the effects of retuning the coils to different load conditions and the influence of a large bore radiofrequency‐shield have been examined. The calculation time for SAR simulations in the head can be reduced by 50% without significant error for smaller model extent and simplified tissue structure outside the coil. Likewise, the model generation can be accelerated by reducing the number of tissue types. Local SAR can vary up to 14% due to position alone. This must be considered and sets a limit for SAR prediction accuracy. All these results are comparable between the two body models tested. Magn Reson Med 69:1157–1168, 2013.Keywords
Funding Information
- BMBF-funded INUMAC project (13N9206)
This publication has 29 references indexed in Scilit:
- MRI-based anatomical model of the human head for specific absorption rate mappingMedical & Biological Engineering & Computing, 2008
- 7 T body MRI:B1shimming with simultaneous SAR reductionPhysics in Medicine & Biology, 2007
- Calculation of SAR for transmit coil arraysConcepts in Magnetic Resonance Part B, 2007
- Calculations ofB 1 distribution, specific energy absorption rate, and intrinsic signal-to-noise ratio for a body-size birdcage coil loaded with different human subjects at 64 and 128 MHzApplied Magnetic Resonance, 2005
- Transmit SENSEMagnetic Resonance in Medicine, 2002
- Signal‐to‐noise ratio and absorbed power as functions of main magnetic field strength, and definition of “90°” RF pulse for the head in the birdcage coilMagnetic Resonance in Medicine, 2001
- SAR and B1 field distributions in a heterogeneous human head model within a birdcage coilMagnetic Resonance in Medicine, 1998
- Computation of electromagnetic fields for high-frequency magnetic resonance imaging applicationsPhysics in Medicine & Biology, 1996
- Spatial distribution of high-frequency electromagnetic energy in human head during MRI: numerical results and measurementsIEEE Transactions on Biomedical Engineering, 1996
- Magnetic resonance imaging: Calculation of rates of energy absorption by a human‐torso modelBioelectromagnetics, 1990