Comparison of the Levitan FPS Scope ™ and the single‐use bougie for simulated difficult intubation in anaesthetised patients
Open Access
- 18 April 2007
- journal article
- research article
- Published by Wiley in Anaesthesia
- Vol. 62 (5) , 509-515
Abstract
Summary A randomised cross‐over study was performed in 34 patients with no evidence of airway difficulties, following induction of general anaesthesia, to compare the efficacy of the Levitan FPS scope ™ (LFPS) and the single‐use bougie for tracheal intubation during simulated grade IIIa laryngoscopy. Success rates for intratracheal placement of the device, and the time required for insertion and tracheal intubation were recorded. Both devices were equally successful (31/34 for the LFPS vs 29/34 for the bougie) for insertion into the glottis. The mean insertion time for the LFPS was statistically longer than that for the bougie (4.4–12.5 s) but this difference was not clinically relevant. Intubation times were similar between the two devices. Major problems hindering successful intubation using the LFPS were the presence of a narrow epiglottic‐pharyngeal wall space and copious secretions. An inability to maintain the desired shape was the principal cause of failure with the bougie.Keywords
This publication has 29 references indexed in Scilit:
- A comparison of the Seeing Optical Stylet and the gum elastic bougie in simulated difficult tracheal intubation: a manikin studyAnaesthesia, 2006
- The Effect of Stylet Bend Angle on Tracheal Tube Passage Using a Straight-to-Cuff ShapeAnnals of Emergency Medicine, 2005
- Tracheal intubation using the Bonfils intubation fibrescope after failed direct laryngoscopy*Anaesthesia, 2004
- A comparison of the intubating laryngeal mask airway and the Bonfils intubation fibrescope in patients with predicted difficult airways*Anaesthesia, 2004
- ‘Hold up’ and the gum elastic bougieAnaesthesia, 2003
- Using the Shikani Optical StyletJournal of Bronchology, 2003
- New “Seeing” Stylet‐Scope and Method for the Management of the Difficult AirwayOtolaryngology–Head and Neck Surgery, 1999
- Simulation technique for difficult intubation: Teaching tool or new hazard?Journal of Clinical Anesthesia, 1990
- Successful difficult intubationAnaesthesia, 1988
- Difficult tracheal intubationAnaesthesia, 1986