Abstract
The pH dependent conductance of the Chitosan/Gelatin and Hydroxy-Ethyl Methacrylate (HEMA)/Gelatin hydrogel based systems was mapped out using a data acquisition system and several aspects of conductance characteristics namely accuracy, response time and sensitivity were compared and discussed. The material characteristics of these sensors were mapped out through FTIR. Specific accuracy tests were designed and conducted and both hydrogel based sensors have been found to give a reasonably predictable outcomes in the physiological pH range (4–10). Although Chitosan/Gelatin sensing systems have been explored with regards to conductivity, to the best of our knowledge, HEMA/Gelatin composite has not yet been explored in this respect. There are observable differences in the conductometric system analysis results of both hydrogels; it has been found that HEMA based hydrogels have a longer response time, but also offer a better sensitivity as compared to Chitosan based conductometric hydrogels. This difference in response may be attributed to the material based characteristics of the conductometric sensing system; the possible underlying mechanisms have also been discussed in this research. Our findings indicate that these conductometric composite hydrogels have the potential to be used in physiological applications; where varying response times are required, a hybrid sensor array comprising of both types of hydrogel sensors may be used; that can utilize the advantages of faster response time in case of Chitosan/Gelatin and better sensitivity in case of HEMA/Gelatin sensors.

This publication has 17 references indexed in Scilit: