The importance of electrode location in cochlear implantation
Authors:
Open Access
- 29 November 2016
- journal article
- review article
- Published by Wiley in Laryngoscope Investigative Otolaryngology
- Vol. 1 (6) , 169-174
Abstract
Objectives: As indications for cochlear implantation have expanded to include patients with more residual hearing, increasing emphasis has been placed on minimally traumatic electrode insertion. Histopathologic evaluation remains the gold standard for evaluation of cochlear trauma, but advances in imaging techniques have allowed clinicians to determine scalar electrode location in vivo. This review will examine the relationship between scalar location of electrode arrays and audiologic outcomes. In addition, the impact that surgical approach, electrode design, and insertion depth have on scalar location will be evaluated.Data Sources: PubMed literature reviewReview Methods: A review of the current literature was conducted to analyze the relationship between scalar location of cochlear implant electrode arrays and speech perception outcomes. Further, data were reviewed to determine the impact that surgical variables have on scalar electrode location.Results: Electrode insertions into the scala tympani are associated with superior speech perception and higher rates of hearing preservation. Lateral wall electrodes, and round window/extended round window approaches appear to maximize the likelihood of a scala tympani insertion. It does not appear that deeper insertions are associated with higher rates of scalar translocation.Conclusion: Superior audiologic outcomes are observed for electrode arrays inserted entirely within the scala tympani. The majority of clinical data demonstrate that lateral wall design and a round window approach increase the likelihood of a scala tympani insertion.Level of Evidence: N/A.Keywords
This publication has 43 references indexed in Scilit:
- Implications of Minimizing Trauma During Conventional Cochlear ImplantationOtology & Neurotology, 2011
- Anatomic verification of a novel method for precise intrascalar localization of cochlear implant electrodes in adult temporal bones using clinically available computed tomographyThe Laryngoscope, 2010
- Psychophysical Properties of Low-Frequency Hearing: Implications for Perceiving Speech and Music via Electric and Acoustic Stimulation2009
- Impact of Low-Frequency HearingAudiology and Neurotology, 2009
- Integration of acoustic and electrical hearingJournal of Rehabilitation Research and Development, 2008
- Role of Electrode Placement as a Contributor to Variability in Cochlear Implant OutcomesOtology & Neurotology, 2008
- Variance of Angular Insertion Depths in Free-Fitting and Perimodiolar Cochlear Implant ElectrodesOtology & Neurotology, 2008
- Scala Tympani Cochleostomy I: Results of a SurveyThe Laryngoscope, 2007
- Quality Control After Insertion of the Nucleus Contour and Contour Advance Electrode in AdultsEar & Hearing, 2007
- Surgical implications of perimodiolar cochlear implant electrode design: avoiding intracochlear damage and scala vestibuli insertionCochlear Implants International, 2001