The Value of Quality Improvement Process in the Detection and Correction of Common Errors in Echocardiographic Hemodynamic Parameters in a Busy Echocardiography Laboratory
Authors:
- 1 June 2015
- journal article
- research article
- Published by Wiley in Echocardiography
- Vol. 32 (12) , 1778-1789
Abstract
Background: Accurate assessment of cardiac structures, ventricular function, and hemodynamics is essential for any echocardiographic laboratory. Quality improvement (QI) processes described by the American Society of Echocardiography (ASE) and the Intersocietal Commission (IAC) should be instrumental in reaching this goal.Methods: All patients undergoing transthoracic echocardiogram (TTE) followed by cardiac catheterization within 24 hours at Christiana Care Health System in 2011 and 2012 were identified, with 126 and 133 cases, respectively. Hemodynamic parameters of diastolic function and pulmonary artery systolic pressure (PASP) on TTE correlated poorly with catheterization in 2011. An educational process was developed and implemented at quarterly QI meetings based on ASE and IAC recommendations to target frequently encountered errors and provide methods for improved performance. The hemodynamic parameters were then reexamined in 2012 postintervention.Results: Following the QI process, there was significant improvement in the correlation between invasive and echocardiographic hemodynamic measurements in both systolic and diastolic function, and PASP. This reflected in significant better correlations between echo and cath LVEF [R = 0.88, ICC = 0.87 vs. R = 0.85, ICC = 0.85; P < 0.001], average E/E′ and of left ventricle end‐diastolic pressure (LVEDP) [R = 0.62 vs. R = 0.09, P = 0.006] and a better correlation for PASP [R = 0.77, ICC = 0.77 vs. R = 0.30, ICC = 0.31; P = 0.05] in 2012 compared to 2011.Conclusion: The QI process, as recommended by ASE and IAC, can allow for identification as well as rectification of quality issues in a large regional academic medical center hospital.Keywords
Funding Information
- Institutional Development Award (IDeA)
- National Institute of General Medical Sciences of the National Institutes of Health (U54-GM104941)
This publication has 10 references indexed in Scilit:
- Improved Interobserver Variability and Accuracy of Echocardiographic Visual Left Ventricular Ejection Fraction Assessment through a Self-Directed Learning Program Using Cardiac Magnetic Resonance ImagesJournal of the American Society of Echocardiography, 2013
- Quality in Echocardiography: Does It Make a Difference?Journal of the American Society of Echocardiography, 2013
- Improvement in the Assessment of Diastolic Function in a Clinical Echocardiography Laboratory Following Implementation of a Quality Improvement InitiativeJournal of the American Society of Echocardiography, 2011
- Can a Teaching Intervention Reduce Interobserver Variability in LVEF AssessmentJACC: Cardiovascular Imaging, 2011
- American Society of Echocardiography Recommendations for Quality Echocardiography Laboratory OperationsJournal of the American Society of Echocardiography, 2011
- Guidelines for the Echocardiographic Assessment of the Right Heart in Adults: A Report from the American Society of EchocardiographyJournal of the American Society of Echocardiography, 2010
- Recommendations for the Evaluation of Left Ventricular Diastolic Function by EchocardiographyJournal of the American Society of Echocardiography, 2009
- Prospective Evaluation of the Clinical Application of the American College of Cardiology Foundation/American Society of Echocardiography Appropriateness Criteria for Transthoracic EchocardiographyJACC: Cardiovascular Imaging, 2008
- Achieving Quality in Cardiovascular ImagingJACC, 2006
- Problems of Spectrum and Bias in Evaluating the Efficacy of Diagnostic TestsThe New England journal of medicine, 1978