An Electrothermal Atomic Absorption Spectrometer Using Semiconductor Diode Lasers and a Tungsten Coil Atomizer: Design and First Applications
Authors:
- 1 August 1998
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 70 (17) , 3525-3532
Abstract
A new type of atomic absorption spectrometer using a laser diode as light source and a tungsten coil as atomizer is described. Compared to established atomic absorption spectrometers, it is much simpler in construction, smaller in size, and less expensive and it provides inherent background correction and high detection power. The performance of this concept is demonstrated by the determination of aluminum and chromium in water, blood serum and, using the slurry sampling technique, in powdered high-purity graphite and titanium dioxide samples. For calibration, the standard addition method was used. Possible interferences by impurities originating from the tungsten coils are discussed. Applying aqueous solutions of Al and Cr, detection limits of 0.9 and 0.03 ng/mL, respectively, were obtained, and for serum, they were 2.5 and 0.3 ng/mL, respectively. For these elements in graphite and titanium dioxide applied as slurry, the detection limits are between 0.02 (Cr in TiO2) and 0.6 μg/g (Al in graphite). The accuracy was checked by comparison of the results with those of other methods. The described system is especially suitable for on-site and on-line analysis.Keywords
This publication has 22 references indexed in Scilit:
- Comparison of electron concentrations, electron temperatures, gas kinetic temperatures, and excitation temperatures in argon ICPs operated at 27 and 40 MHzSpectrochimica Acta Part B: Atomic Spectroscopy, 1997
- Separation and preconcentration by flow injection coupled to tungsten coil electrothermal atomic absorption spectrometrySpectrochimica Acta Part B: Atomic Spectroscopy, 1996
- Slurry sampling electrothermal atomic absorption spectrometry for the analysis of graphite powdersSpectrochimica Acta Part B: Atomic Spectroscopy, 1996
- Peer Reviewed: Semiconductor Diode Lasers in Atomic SpectrometryAnalytical Chemistry, 1996
- Portable, Battery-Powered, Tungsten Coil Atomic Absorption Spectrometer for Lead DeterminationsApplied Spectroscopy, 1996
- Analytical evaluation of a tungsten coil atomizer for cadmium, lead, chromium, manganese, nickel and cobalt determination by electrothermal atomic absorption spectrometryAnalytica Chimica Acta, 1995
- Electrothermal atomic absorption spectrometric determination of Al, Cu, Fe, Pb, V and Zn in clinical samples and in certified environmental reference materialsAnalytica Chimica Acta, 1994
- Determination of chromium in serum and lake water by electrothermal atomic absorption spectrometry using vanadium and molybdenum modifierJournal of Analytical Atomic Spectrometry, 1994
- Simple low-cost tungsten-coil atomiser for electrothermal atomic absorption spectrometryJournal of Analytical Atomic Spectrometry, 1988
- The application of electrodeposition techniques to flameless atomic absorption spectrometryAnalytica Chimica Acta, 1974