Intercalary reconstruction of long bones by massive allograft: Comparison of construct stability ensured by three different host-graft junctions and two types of fixations in a synthetic femur model
Open Access
- 3 August 2022
- journal article
- research article
- Published by Frontiers Media SA in Frontiers in Pediatrics
- Vol. 10, 868299
Abstract
An intercalary segmental allograft is an option for limb salvage in bone tumours. Stable and congruent intercalary reconstructions are a prerequisite for achieving host-graft union. However, a too rigid fixation could increase the risk of late complications correlated with negative bone remodelling. This study compared the reconstruction stiffness achieved by three different host-graft junctions, namely, end-to-end, modified step-cut, and taper. A low-stiffness bone plate was used as the fixation method, except for the taper junction where a low-stiffness intramedullary nail was also used to investigate the effects of different types of fixation on construct stiffness. Composite femora were tested under four loading conditions to determine coronal and sagittal bending stiffness, as well as torsional stiffness in opposite directions. Stiffness values were expressed as a percentage of intact host bone stiffness (%IBS). While a reduction of coronal bending stiffness was found with taper junctions (76%IBS) compared with the high values ensured by end-to-end (96%IBS) and modified step-cut junctions (92%IBS), taper junctions significantly increased stiffness under sagittal bending and torsion in intra- and extra-direction: end-to-end 29%IBS, 7%IBS, 7%IBS, modified step-cut 38%IBS, 20%IBS, 21%IBS, and taper junction 52%IBS, 55%IBS, 56%IBS, respectively. Construct stiffness with taper junctions was decreased by 11–41%IBS by replacing the bone plate with an intramedullary nail. Taper junctions can be an alternative to achieve intercalary reconstructions with more homogeneous and, in three out of four loading conditions, significantly higher construct stability without increasing bone plate stiffness. The risk of instability under high torsional loads increases when taper junctions are associated with a low-stiffness intramedullary nail.Keywords
This publication has 54 references indexed in Scilit:
- Patient-related risk factors for fracture-healing complications in the United Kingdom General Practice Research DatabaseActa Orthopaedica, 2012
- Comparison of mechanical rigidity between plate augmentation leaving the nail in situ and interlocking nail using cadaveric fracture model of the femurInternational Orthopaedics, 2010
- Structural behaviour and strain distribution of the long bones of the human lower limbsJournal of Biomechanics, 2010
- Telescope Allograft Method to Reconstitute the Diaphysis in Limb Salvage SurgeryClinical Orthopaedics and Related Research, 2009
- Survival of massive allografts in segmental oncological bone defect reconstructionsInternational Orthopaedics, 2008
- Combined Massive Allograft and Intramedullary Vascularized Fibula Transfer: The Capanna Technique for Lower-Limb ReconstructionSeminars in Plastic Surgery, 2008
- Artificial composite bone as a model of human trabecular bone: The implant–bone interfaceJournal of Biomechanics, 2007
- A technique for enhancing union of allograft to host boneThe Journal of Bone and Joint Surgery. British volume, 1998
- Cross‐sectional geometry of Pecos Pueblo femora and tibiae—A biomechanical investigation: II. Sex, age, and side differencesAmerican Journal of Physical Anthropology, 1983
- Cross‐sectional geometry of Pecos Pueblo femora and tibiae—A biomechanical investigation: I. Method and general patterns of variationAmerican Journal of Physical Anthropology, 1983