Clinical validation of a continuous non-invasive haemodynamic monitor (CNAP™ 500) during general anaesthesia
Authors:
Open Access
- 3 February 2012
- journal article
- research article
- Published by Elsevier in British Journal of Anaesthesia
- Vol. 108 (4) , 581-585
Abstract
We conducted a prospective study to test the validity of a new continuous non-invasive blood pressure (NIBP) monitor (CNAP) (CNAP™ Monitor 500). One hundred patients undergoing elective surgery under general anaesthesia were included in the study after informed written consent. The CNAP finger cuffs were placed on the fingers of one arm, an arterial catheter was inserted into the same arm and data were recorded simultaneously. Agreement between invasive arterial pressure (IAP) and blood pressure obtained by CNAP was compared using the Bland–Altman method for repeated measurements. The data from the first 50 patients (software V3.0) were used to improve the software of the CNAP (software V3.5), which was then evaluated in another 50 patients. We defined a clinically acceptable agreement according to the standards of the American Association for the Advancement of Medical Instrumentation for NIBP measurements [limits of agreement (LOA) ±15 mm Hg]. We analysed 524 878 paired measurements in 100 patients. The mean bias of the mean arterial pressure in the first 50 patients was −2.9 mm Hg (sd 10.6 mm Hg, LOA −23.7 to 17.9 mm Hg), and in the consecutive 50 patients (using software V3.5) the bias was −3.1 mm Hg (sd 9.5 mm Hg, LOA −21.6 to 15.4 mm Hg). The new CNAP monitor showed an agreement with the IAP that is promising but did not match our predefined criteria.Keywords
This publication has 13 references indexed in Scilit:
- BioSig: The Free and Open Source Software Library for Biomedical Signal ProcessingComputational Intelligence and Neuroscience, 2011
- Precision and accuracy of a new device (CNAP™) for continuous non-invasive arterial pressure monitoring: assessment during general anaesthesiaBritish Journal of Anaesthesia, 2010
- Continuous Non-Invasive Arterial Pressure Technique Improves Patient Monitoring during Interventional EndoscopyInternational Journal of Medical Sciences, 2009
- I. Using the Bland–Altman method to measure agreement with repeated measuresBritish Journal of Anaesthesia, 2007
- Continuous non-invasive blood pressure monitoring using concentrically interlocking control loopsComputers in Biology and Medicine, 2006
- Arterial catheter-related infection of 2,949 cathetersCritical Care, 2006
- Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adultsBlood Pressure Monitoring, 2002
- Clinical review: Complications and risk factors of peripheral arterial catheters used for haemodynamic monitoring in anaesthesia and intensive care medicineCritical Care, 2002
- Measuring agreement in method comparison studiesStatistical Methods in Medical Research, 1999
- Fifteen years experience with finger arterial pressure monitoring: assessment of the technologyCardiovascular Research, 1998