Abstract
Bone density at the implant site is still assessed subjectively, with high inter- and intra-observer variability. This ex vivo study evaluated the repeatability of electrochemical impedance spectroscopy (EIS) of bone and its correlation with implant primary stability, measured by resonance frequency analysis (ISQ). Thirty bovine bone samples, each from a different animal, underwent triplicate EIS scans (3–99.1 kHz) within 6 h post-mortem; the Quality Bone Parameter (QBP) was defined as the base-10 logarithm of the impedance modulus at 99.1 kHz. A 3.8 × 10 mm implant was then placed after standardized drilling, and ISQ was measured immediately. Two samples with non-physical acquisitions (negative real impedance) were excluded. In the remaining 28 samples, QBP was highly repeatable (median coefficient of variation 0.22%) and showed a very strong correlation with ISQ (Pearson r = 0.98, 95% CI 0.95–0.99, p < 0.001; Spearman ρ = 0.97), which strengthened with increasing frequency. Under these standardized ex vivo conditions, high-frequency bone impedance was therefore closely associated with ISQ. These findings cannot be extrapolated directly to clinical settings, where the repeatability of the measurement and the influence of tissue hydration remain to be investigated in vivo.