Association Between an Electrochemical Impedance Parameter of Bone and Dental Implant Primary Stability: An Ex Vivo Study on Bovine Bone
Preprint
- 20 October 2026
- preprint
- Published by MDPI AG in Preprints.org
Abstract
Bone density at the implant site is still assessed subjectively, with high inter- and intra-observer variability. This ex vivo study evaluated the repeatability of electrochemical impedance spectroscopy (EIS) of bone and its correlation with implant primary stability, measured by resonance frequency analysis (ISQ). Thirty bovine bone samples, each from a different animal, underwent triplicate EIS scans (3–99.1 kHz) within 6 h post-mortem; the Quality Bone Parameter (QBP) was defined as the base-10 logarithm of the impedance modulus at 99.1 kHz. A 3.8 × 10 mm implant was then placed after standardized drilling, and ISQ was measured immediately. Two samples with non-physical acquisitions (negative real impedance) were excluded. In the remaining 28 samples, QBP was highly repeatable (median coefficient of variation 0.22%) and showed a very strong correlation with ISQ (Pearson r = 0.98, 95% CI 0.95–0.99, p < 0.001; Spearman ρ = 0.97), which strengthened with increasing frequency. Under these standardized ex vivo conditions, high-frequency bone impedance was therefore closely associated with ISQ. These findings cannot be extrapolated directly to clinical settings, where the repeatability of the measurement and the influence of tissue hydration remain to be investigated in vivo.Keywords
All Related Versions
This publication has 26 references indexed in Scilit:
- Comprehensive characterization of osseous tissues from impedance measurements by effective medium approximationAIP Advances, 2021
- Array programming with NumPyNature, 2020
- SciPy 1.0: fundamental algorithms for scientific computing in PythonNature Methods, 2020
- Potential of electrical impedance spectroscopy to differentiate between healthy and osteopenic boneClinical Biomechanics, 2018
- Characterization of the electrical conductivity of bone and its correlation to osseous structureScientific Reports, 2018
- Interrelationships between electrical, mechanical and hydration properties of cortical boneJournal of the Mechanical Behavior of Biomedical Materials, 2018
- New opportunities for fracture healing detection: Impedance spectroscopy measurements correlate to tissue composition in fracturesJournal of Orthopaedic Research, 2017
- Micro-architecture and mineralization of the human alveolar bone obtained with microCTArchives of Oral Biology, 2013
- Timing of loading and effect of micromotion on bone-dental implant interface: Review of experimental literatureJournal of Biomedical Materials Research, 1998
- Quantitative determination of the stability of the implant‐tissue interface using resonance frequency analysisClinical Oral Implants Research, 1996